Flow Rate Settings
1 to 6 setting in Pulse Flow Mode
0.5 to 3.0 LPM in Continuous Flow Mode
Oxygen Concentration
90% +/– 3% for all flow settings
Oxygen Concentration Indicator
Green Light = Normal Operation
Yellow Light = Warning or Caution, less than 85% +/– 3%
Red Light Flashing = Abnormal Operation, less than 70% +/– 5%
Maximum System Pressure
15 psig (103.5 kPa)
Oxygen Outlet Pressure
5.0 psig (34.5 kPa) nominal
Electrical Power:
AC Operation
100 – 240 VAC, 50/60 Hz
DC Operation
12 V
Power Cartridge
195 W-Hrs
AC Power Supply
200 Watts Max
DC Power Supply
150 Watts Max
Operating Temperature
50° F to 104° F (10° C to 40° C)
Operating Humidity
10% - 95% at 82.4° F (28° C) dew point
Nominal Sound Level
48 dBA at 3.0 LPM in Continuous Flow Mode
40 dBA at 3.0 LPM in Pulse Flow Mode
Dimensions (H x W x D)
19.3 inches x 12.3 inches x 7.1 inches
(49.0 cm x 31.2 cm x 18.0 cm)
Weight
17.9 pounds (8.1 kg) with Power Cartridge
14.5 pounds (6.6 kg) without Power Cartridge
Continuous Flow Indication
Expressed in liters per minute (LPM)
Audible Alarm Indicators
Loss of power/Hot Power Cartridge
Low Battery/Warm Power Cartridge
Low Oxygen Output
O
2
flow outside normal limits
No inspiration detected in Pulse Mode
Unit Malfunction
Back-Up Alarm Power
9V Internal Battery
Filters
Air Inlet, HEPA
Device Classification
IEC Class I, Type B Applied Part, IPX0
Transport/Storage Requirements
Temperature: -4° F (-20°C) to +140° F (+60° C)
Humidity: up to 95% Non-Condensing
Back pressure Effect on Flow Indicator
None
Altitude
0 –13,123 feet (0 – 4,000 meters)
Technical Data
*
38
|
ECLIPSE
TM
Model 1000
User Manual
StockCheck.com