background image

Operating Your ECLIPSE

STEP 1:

LOCATING YOUR ECLIPSE

Place the ECLIPSE in a well-ventilated place so that the air inlet and exhaust 
have clear access.

Position the ECLIPSE so that any alarm can be heard.

Check to see if the air inlet filter is in place before using. If the air inlet filter needs
to be cleaned, wash with soapy water and allow drying before using. If necessary,
replace with a new air inlet filter.

Plug unit into AC Power, DC Power, or make sure there is a charged Power
Cartridge installed. 

Model 1000

User Manual

ECLIPSE

TM

Model 1000   

User Manual

|

25

StockCheck.com

Содержание RatioMatic 1000

Страница 1: ...Model 1000User Manual S t o c k C h e c k c o m...

Страница 2: ...S t o c k C h e c k c o m...

Страница 3: ...Panel Features 12 User Controls 13 Front Panel Icons 13 System Status Indicators 13 Recommended Operating Environments 15 The Proper Location 16 Operating Instructions 17 Before Operating 17 Powering...

Страница 4: ...ying information specific instructions commentary sidelights or interesting points Symbols Used in User Manual and on Oxygen System Refer to this section for an explanation of the symbols and the warn...

Страница 5: ...ant information about the possibility of spontaneous and violent ignition that may occur if oil grease or other petroleum based substances come into contact with oxygen under pressure Drip Proof Equip...

Страница 6: ...ng Decreases the flow setting in 0 5 increments each time the button is pushed Flow Mode Activates either Continuous Mode or Pulse Flow Mode Operation External Power Present Indicator Indicates the pr...

Страница 7: ...NG CAPABILITIES WARNING IN CERTAIN CIRCUMSTANCES THE USE OF NON PRESCRIBED OXYGEN CAN BE HAZARDOUS THIS DEVICE SHOULD ONLY BE USED WHEN PRESCRIBED BY A PHYSICIAN WARNING FOR ADULT USE ONLY NOT QUALIFI...

Страница 8: ...just the flow only to your prescribed flow rate unless otherwise instructed by your physician WARNING No Smoking or Naked Flames The use of oxygen therapy requires that special care be taken to reduce...

Страница 9: ...your oxygen system has been dropped damaged or exposed to water contact your dealer provider for examination or repair DO NOT drop or insert any object into any opening DO NOT block the air inlet or t...

Страница 10: ...SeQual AC Power Supply and DC Power Supply supplied with your ECLIPSE Oxygen System Use of a non SeQual Power Supply may be hazardous cause damage to your oxygen system and will void the warranty CAU...

Страница 11: ...e Power Cartridge away from children WARNING Use of a damaged Power Cartridge may cause personal injury WARNING Exposing the Power Cartridge to water or other liquids may cause personal injury CAUTION...

Страница 12: ...T FOR CHRONIC HYPOXIA CAUSED BY CHRONIC OBSTRUCTIVE PULMONARY DISEASE OXYGEN THERAPY HAS BECOME MORE VERSATILE IN RECENT YEARS DUE TO THE DEVELOPMENT OF OXYGEN CONCENTRATORS THE ECLIPSE MAKES IT EASIE...

Страница 13: ...ust and debris from entering the ECLIPSE and should be cleaned regularly Service Port This connection allows service personnel to diagnose and monitor ECLIPSE performance It is not for patient use Ext...

Страница 14: ...he Controls and Front Panel continued 12 ECLIPSETM Model 1000 User Manual 0 0 0 6 No Smoking Icon ON OFF Button ON OFF Indicator Green Alert Indicator Yellow Flow Rate Setting 2 Digits Green Alarm Ind...

Страница 15: ...is delivered upon inhalation When the pulse of oxygen is delivered the indicator briefly goes off Only change your flow settings at the recommendation of your physician SYSTE M STATUS I N D ICATOR S...

Страница 16: ...lse Flow settings if pulse flow has been prescribed DO NOT change these settings unless it is under the order of your physician Each time you turn the unit on the last flow mode and flow setting will...

Страница 17: ...s draperies furniture etc Environment Smoke pollutant and fume free Operating Time Up to 24 hours per day when connected to external power When moving the ECLIPSE from an extreme environment allow tim...

Страница 18: ...se without being restricted Locate the unit such that the alarms can be heard Route oxygen tubing such that it does not kink or is subject to occlusion DO NOT cover unit with a blanket towel quilt or...

Страница 19: ...ection of this manual Follow all operating instructions Observe all warnings on the product and in the operating instructions To reduce the risk of fire bodily injuries and damage to the oxygen system...

Страница 20: ...power is restored by plugging into AC or DC outlets the system will accept power from the external source and replenish the Power Cartridge while operating if adequate power is available Connect the...

Страница 21: ...eceptacle DO NOT remove ground pin from cord Use only with SeQual supplied cords CAUTION DO NOT connect the ECLIPSE to an extension cord or electrical outlet controlled by a switch 2 DC POWER SUPPLY A...

Страница 22: ...y from the vehicle for low voltage alarm conditions WARNING If the vehicle power source vehicle accessory connector drops below 11 5 Volts brownout the ECLIPSE will revert to Power Cartridge operation...

Страница 23: ...stem When the Power Cartridge is properly inserted and the ECLIPSE is ON the Power Cartridge Status Gauge will appear on the Control Panel If you DO NOT have a Power Cartridge installed the Status Gau...

Страница 24: ...UTION DOT UN Regulations require the removal of the Power Cartridge for international travel CAUTION There are no serviceable parts inside the Power Cartridge DO NOT open CAUTION Replace the Power Car...

Страница 25: ...artridge packaged with your ECLIPSE Oxygen System is not fully charged Before using your ECLIPSE Oxygen System for the first time you must fully discharge the Power Cartridge 1 Turn your ECLIPSE ON an...

Страница 26: ...artridge Use of another Power Cartridge or Battery may damage the unit present a risk of fire or explosion and will void the warranty In the event of a power interruption the ECLIPSE Oxygen System wil...

Страница 27: ...be heard Check to see if the air inlet filter is in place before using If the air inlet filter needs to be cleaned wash with soapy water and allow drying before using If necessary replace with a new...

Страница 28: ...nternal malfunction or power outage external power loss with no Power Cartridge installed The red yellow and green all OK indicators will initially light up and then the red and yellow indicators sequ...

Страница 29: ...r cannula and tubing cleaning and replacement information STEP 4 SELECT FLOW MODE Press the Flow Mode button to select the flow delivery mode Continuous Flow Mode Operation When operating in Continuou...

Страница 30: ...Settings from Continuous Flow applications may not be applicable to Pulse Flow Mode WARNING When operating in Pulse Flow Mode to ensure adequate breath detection the length of the tubing cannula must...

Страница 31: ...ator holes on the bottom of the unit to the tabs on the cart Align the screw to the insert on the back of the ECLIPSE and hand tighten the knob on the cart Push the button on the cart handle to adjust...

Страница 32: ...Off Blinking Beeps Continue Use Check or clean air intake filter If condition persists contact dealer provider Flow Rate Error Off Blinking Off Beeps Check tubing and or humidifier Blocked Flow for ob...

Страница 33: ...nal power Malfunction Contact your dealer provider Low Oxygen Concentration Restriction in tubing or humidifier Repair or replace tubing or humidifier Filter restricted Clean or replace air inlet filt...

Страница 34: ...t port of the oxygen system The O2 symbol marks the oxygen outlet port on the concentrator Use of humidifiers not specified for this oxygen system may impair performance and void the warranty DO NOT u...

Страница 35: ...ibed a cannula to deliver oxygen In most cases they are already attached to the oxygen tubing If not follow the instructions included with the cannula to attach it to the oxygen tubing Use of cannula...

Страница 36: ...een or red Green indicates normal operation and red indicates abnormal operation WARNING DO NOT expose the AC and DC Power Supplies to water Electrical shock or serious injury may result Cart Your ECL...

Страница 37: ...your dealer provider for any maintenance required Patient Cleaning and Care Cannula Replacement Replace your tubing and cannula on a regular basis as recommended by your dealer provider Your physicia...

Страница 38: ...rn OFF the ECLIPSE and disconnect from AC or DC power before any cleaning or disinfection activity Use a damp not soaking wet cloth or sponge and a mild detergent solution to clean the cabinet and pow...

Страница 39: ...charge to calibrate the Power Cartridge Use a damp cloth or sponge and mild detergent to clean the Power Cartridge case and latch WARNING DO NOT tamper with or try to repair the Power Cartridge No ser...

Страница 40: ...LPM in Continuous Flow Mode 40 dBA at 3 0 LPM in Pulse Flow Mode Dimensions H x W x D 19 3 inches x 12 3 inches x 7 1 inches 49 0 cm x 31 2 cm x 18 0 cm Weight 17 9 pounds 8 1 kg with Power Cartridge...

Страница 41: ...al a pulse of oxygen is released to the patient by an electronically controlled flow valve There is no mixing of oxygen with room air Range of Operation Settings Increments of 1 6 times x the bolus of...

Страница 42: ...RED DEVICE THE USER MAY EXPERIENCE PERIODS OF NON OPERATION AS THE RESULT OF ELECTRICAL POWER INTERRUPTION OR THE NEED TO HAVE THE OXYGEN CONCENTRATOR SERVICED BY A TECHNICIAN AN OXYGEN CONCENTRATOR I...

Страница 43: ...he unit may result WARNING Protect electrical cords from sharp edges Electrical shock and serious physical injury may result WARNING Use only the SeQual AC Power Supply and DC Power Supply supplied wi...

Страница 44: ...ure the ECLIPSE and the DC Power Supply in your vehicle and make sure the air inlet and exhaust airflow are not blocked This may impact performance WARNING DO NOT leave the ECLIPSE or DC Power Supply...

Страница 45: ...may not be able to detect some respiratory efforts WARNING DO NOT change from prescribed flow rate setting without consulting with a qualified clinician WARNING AVAILABILITY OF AN ALTERNATE SOURCE OF...

Страница 46: ...low Disassembling the oxygen system may create a hazard of electrical shock and void your warranty Contact your dealer provider for any maintenance required WARNING Unplug Power Cords AC or DC Power S...

Страница 47: ...C such as in a vehicle parked in the sun or on a hot day CAUTION Charging the Power Cartridge below 50 F 10 C or above 104 F 40 C may degrade the performance CAUTION DO NOT drop the Power Cartridge o...

Страница 48: ...ll help to assure the longevity of your Power Cartridge CAUTION DOT UN Regulations require the removal of the Power Cartridge for international travel CAUTION There are no serviceable parts inside the...

Страница 49: ...e tubing must not exceed 50 feet 15 2m when using Continuous Flow Mode and must not exceed 7 feet 2 1m when operating in Pulse Flow Mode for adequate breath detection CAUTION When operating the ECLIPS...

Страница 50: ...icker emissions IEC 61000 3 3 This equipment has been tested and found to comply with the limits for medical devices to the IEC60601 1 2 2001 09 Electromagnetic Compatibility standard These limits are...

Страница 51: ...Sorrento Valley Road San Diego CA 92121 USA 858 202 3100 800 826 4610 www sequal com EUROPEAN UNION AUTHORIZED AGENT MeddiQuest Limited Business Technology Centre Bessemer Drive Stevenage SG1 2DX Unit...

Страница 52: ...ECLIPSETM Model 1000 User Manual PN 6112 03 06 S t o c k C h e c k c o m...

Отзывы: