background image

 

 

FCA

22

 

Instruction

 

Manual

  

 

 

 

 

Page 5 

 

 

 

 

Influences

 

on

 

the

 

Measurement

 

pH

 

Value

 

The

 

FCS

 

only

 

measures

 

the

 

HOCl

 

component

 

of

 

the

 

total

 

Free

 

Chlorine.

 

The

 

HOCl

 

proportion

 

varies

 

from

 

100%

 

at

 

pH

 

5.5

 

to

 

0%

 

at

 

pH

 

10,

 

see

 

Figure

 

2.1.

 

The

 

PHS10

 

pH

 

sensor

 

provides

 

automatic

 

compensation

 

for

 

the

 

pH

 

dependent

 

ratio

 

of

 

HOCl

 

and

 

OCl

 

present

 

in

 

the

 

water.

  

The

 

PTFE

 

membrane

 

rejects

 

charged

 

ions

 

allowing

 

only

 

neutral

 

molecules

 

to

 

pass

 

through.

 

Salts

 

and

 

other

 

ionic

 

substances

 

are

 

blocked

 

by

 

the

 

membrane

 

eliminating

 

any

 

influence

 

on

 

the

 

measurement

 

by

 

changes

 

in

 

the

 

conductivity

 

of

 

the

 

sample.

 

The

 

HOCl

 

portion

 

of

 

the

 

free

 

chlorine

 

passes

 

through

 

the

 

membrane

 

to

 

the

 

cathode

 

and

 

is

 

measured

 

while

 

the

 

negatively

 

charged

 

hypochlorite

 

ion,

 

OCl

,

 

portion

 

is

 

rejected

 

by

 

the

 

membrane.

 

The

 

C

22

 

analyzer

 

uses

 

the

 

pH

 

from

 

the

 

PHS10

 

sensor

 

to

 

calculate

 

the

 

OCl

 

value.

 

The

 

measured

 

value

 

and

 

the

 

calculated

 

value

 

are

 

combined

 

and

 

displayed

 

as

 

Free

 

Chlorine.

   

The

 

DPD

 

calibration

 

method

 

measures

 

the

 

Total

 

Chlorine

 

by

 

buffering

 

the

 

sample

 

to

 

pH

 

6.3,

 

thereby

 

converting

 

the

 

entire

 

amount

 

of

 

chlorine

 

to

 

HOCl

 

and

 

then

 

measuring

 

this

 

HOCl

 

component.

 

The

 

pH

 

compensation

 

algorithm

 

in

 

the

 

C

22

 

is

 

designed

 

to

 

match

 

this

 

method.

 

The

 

highest

 

accuracy

 

is

 

attained

 

with

 

calibrations

 

performed

 

at

 

neutral

 

pH

 

values

 

and

 

higher

 

chlorine

 

concentrations.

 

The

 

accuracy

 

of

 

the

 

compensation

 

decreases

 

when

 

the

 

pH

 

is

 

above

 

pH

 

8

 

since

 

there

 

is

 

little

 

chlorine

 

to

 

measure

 

and

 

a

 

large

 

compensation

 

to

 

perform.

  

 

 

Figure

 

2.1

 

Flow

 

The

 

FCS

 

consumes

 

chlorine

 

to

 

produce

 

the

 

signal.

 

The

 

area

 

near

 

the

 

sensing

 

tip

 

will

 

become

 

depleted

 

of

 

chlorine

 

without

 

adequate

 

flow

 

to

 

replenish

 

the

 

sample.

  

The

 

sensor

 

requires

 

a

 

minimum

 

velocity

 

of

 

0.5

 

ft/sec

 

past

 

the

 

membrane.

   

Below

 

this

 

value

 

the

 

sensor

 

will

 

indicate

 

a

 

lower

 

concentration

 

than

 

the

 

actual

 

value.

  

Higher

 

flow

 

rates

 

have

 

little

 

to

 

no

 

effect

 

on

 

the

 

measurement.

 

See

 

Figure

 

2.2.

 

0

20

40

60

80

100

4

6

8

10

%

pH

pH

 

Effects

%

 

HOCl

%

 

OCl

Содержание FCA-22

Страница 1: ...sed in this document may be used and disseminated only for the purposes and to the extent specifically authorized in writing by Electro Chemical Devices Electro Chemical Devices reserves the right to...

Страница 2: ...ample requirements 7 Mounting and Plumbing 7 Powering the C 22 7 Installing the Sensors 7 Start Up 8 Calibration Temperature Sensor 8 pH Sensor 9 Chlorine Sensor 10 Commissioning Outputs 4 20 mA outpu...

Страница 3: ...ow Cell pH Sensor in Flow Cell C 22 Analyzer Controller Part Model and Product Description 1290020 1 Free Chlorine Analyzer with automatic pH compensation complete assembly panel mounted with C 22 tra...

Страница 4: ...mperometric sensor with a PTFE membrane gold cathode and a silver silver chloride anode The C22 analyzer applies a fixed voltage across the electrodes and measures the current flow Hypochlorous acid H...

Страница 5: ...and the calculated value are combined and displayed as Free Chlorine The DPD calibration method measures the Total Chlorine by buffering the sample to pH 6 3 thereby converting the entire amount of ch...

Страница 6: ...ture drops the output drops The FCA 22 automatically compensates for the changes The temperature sensor is located in the pH sensor and it has a response time of several minutes Rapid changes of tempe...

Страница 7: ...o open atmosphere Do not restrict the drain line The sample can be introduced after the sensors have been calibrated and installed in the flow cells Powering the C 22 Attach power as described in the...

Страница 8: ...e agrees with the actual temperature If the value deviates by more than 2 C then adjust the Temperature Trim in the Configuration Trim Menu to the correct value For a detailed description see Temperat...

Страница 9: ...then accepting the calibration For a detailed description see Electrode Standardization in the C 22 manual pH Calibration Screen Shots Calibration 2 point Zero pt and Slope 1 Home Screen Press Menu Se...

Страница 10: ...seconds Press Menu Select Down button to accept the Calibration 13 Return to the Home Screen Leave the Calibration Screen and return to the Home Screen by pressing the CALIBRATE down key Chlorine Cali...

Страница 11: ...be 0 00 ppm Cl2 and the Cal value will display the new mV signal from the sensor Chlorine Calibration Screen Shots 1 Home Screen Press Menu Select Down button 2 Graphical Display Press Menu Select Dow...

Страница 12: ...o the Home Screen Leave the Calibration Screen and return to the Home Screen by pressing the CALIBRATE down key Commissioning Outputs 4 20 mA Range The FCA 22 4 20 mA output is configured 0 20 ppm Fre...

Страница 13: ...A value Factory setting is 0 00 ppm Any value can be entered using the CALIBRATE keys Press the Menu Select down key to accept the value 9 Output Range 20 mA value Press Menu Select Down button 10 Adj...

Страница 14: ...d configuring the relays Screen Shots for Assigning the Alarm Relays 1 Home Screen Press Menu Select Down button 2 Graphical Display Press Menu Select Down button 3 Selection Menu Buffer Press Menu Se...

Страница 15: ...ccept the change press Menu Select down button 15 Select Relay 2 Press Menu Select Down button 16 Access Relay 2 menu To access the menu Simultaneously press both Horizontal CALIBRATE keys 17 Assign R...

Страница 16: ...CALIBRATE keys 5 Choose High Low Alarm Press the horizontal CALIBRATE key to move the cursor under the 6 Set High Low Alarm Press the down CALIBRATE key to change to Notice the Hysteresis sign also c...

Страница 17: ...keys to adjust the value 16 Accept Hysteresis Value Press Menu Select Down button 17 Choose Relay 2 Press Menu Select Down button 18 Set High alarm Simultaneously press both Horizontal CALIBRATE keys...

Страница 18: ...nd at every membrane change Unscrew the measuring chamber from the sensor and pour out the spent solution Fill the chamber to approximately 1 cm from the top and tap it gently to dislodge any trapped...

Страница 19: ...efore returning the sensor to service Constant Head Flow Controller CHFC The CHFC is designed to provide a constant flow to the Free Chlorine Sensor FCS independent of variations in the sample pressur...

Страница 20: ...n DPD test value Insufficient Polarization time Damaged Membrane Cap pH indication higher than actual value Temperature indication lower than actual value Electrical short or wet connection inside the...

Страница 21: ...itting and lift sensor slightly to purge air trapped in the flow cell Refill sensor see Maintenance Calibrate pH sensor or adjust value on uncompensated FCA Calibrate the Temperature see Calibration T...

Страница 22: ...acteristics Accuracy Dependent on the accuracy of the DPD test and pH value Response Time T90 Chlorine 2 minutes Temp 2 minutes pH 30 seconds Drift 1 5 month Polarization time First polarization 60 mi...

Страница 23: ...t incurring any obligation to make any changes in previously delivered products Seller s sole liabilities and the buyer s sole remedies under this agreement shall be limited to a refund in the purchas...

Страница 24: ...zation RMA number must be obtained from the service department before returning any material to ECD All material returned to ECD shall be shipped prepaid to the factory Ship to Electro Chemical Device...

Отзывы: