3.0 OPERATION
3.1
T80
T
RANSMITTER
The ECD Model T80 transmitter is an intelligent, single or dual channel
multi-parameter transmitter designed for the online continuous
measurement of pH, ORP, pION, conductivity, resistivity or Dissolved
Oxygen in a general purpose industrial environment. The Model T80
transmitter digitally communicates with any ECD S80 digital sensor.
The measurement identity is contained in the sensor’s memory. When
an S80 sensor is connected to the transmitter, it automatically
configures the transmitter’s menus an
d display screens to the
measured parameter.
The functions associated with each key are displayed on the screen, above the key for the Selection Adjustment Keys
and to the left of the key for the HOME and BACK keys.
Press any Selection Adjustment key twice within one second
to enter the HOME Menu Screen.
3.2
LQ800
C
ONTROLLER
The ECD Model LQ800 transmitter is an intelligent, single to
eight channel multi-parameter controller designed for the
online continuous measurement of pH, ORP, pION,
conductivity, Resistivity, Dissolved Oxygen, Flow, and Level in a
general purpose industrial environment.
The model LQ800 controller digitally communicates with any
ECD DS80 digital sensor. The measurement identity is contained
in the sensor’s memory. When a DS80 sensor is
connected to
the controller, it automatically configures the controllers,
menus, and display screens to the measured parameter.
3.2.1
S
CREENS
The LQ800 has a touch screen interface that is easy to use. To access Sensor menu screen, tap or press on the Channel
Box of the sensor you would like to access. To access the LQ800 Controller menus and functions, tap or press on the
MENU on the screen.
The
Home
Screen
displays the Controllers Menu button, channels, measurement type, numerical value, the engineering
units, alarms, date and time.
3.2.2
M
ENU
S
TRUCTURE
Pressing the
Menu
button on the home screen will prompt you to the
Menu
Home
Screen. You will be able to set the
analyzer on Hold, and access the
Alarm
logs
,
CONFIGuration
settings, and Info Screens.
3.2.2.1
HOLD
(O
UTPUT
H
OLD
)
Pressing the
Menu
button on the screen will prompt you.
Freezes the 4-20 mA output at the last value prior to activation
Freezes optional Alarm Relays in the current state
Will let you set up the Timeout of the hold system which is configurable with the number entry screen
Pressing HOLD again turns the hold function off, Hold is OFF, displayed. The HOLD function remains ON until it is turned
OFF.
Содержание DO82
Страница 1: ...HYDRA DS NO3 Rev A INSTRUCTIONAL MANUAL HYDRA DS NITRATE ANALYZER...
Страница 18: ...HYDRA DS NO3 N Page 17 Panel Mounting Cut Out 5 35 x 13 6cm x 13 6cm...
Страница 20: ...HYDRA DS NO3 N Page 19 2 4 WIRING 2 4 1 HYDRA DS TO T80 TRANSMITTER...
Страница 41: ...HYDRA DS NO3 N Page 40 Cut Out 5 35 x 5 35 13 6 cm x 13 6 cm 7 2 HYDRA DS SENSOR DIMENSIONAL DRAWING...
Страница 52: ...1500 North Kellogg Drive Anaheim CA 92807 1 714 695 0051 FAX 1 714 695 0057 ecdi com...