![ECD DO82 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/ecd/do82/do82_instructional-manual_3867924014.webp)
HYDRA-DS NO
3
-N
Page 13
1.4
T80
T
RANSMITTER
S
PECIFICATIONS
The ECD Model T80 transmitter is an intelligent single or dual channel multi-parameter transmitter designed for the
online continuous measurement of pH, ORP, pION, conductivity, resistivity, or Dissolved Oxygen in a general purpose
industrial environment. The Model T80 transmitter digitally communicates with any ECD S80 digital sensor. The
measurement identity is contained in the senso
r’s memory. Then an S80 sensor is connected to the transmitter it
automatically configures the transmitter’s menus and display screens to the measured parameter.
The Model T80 transmitter can be loop powered, 24 VDC powered, or 100-240 VAC line powered. The standard
configuration has a 4-20 mA output and an RS485 serial communication port with MODBUS®RTU output. A HART®
communication version (single channel version only) is also available. Alarm relays are optionally available on either
line powered transmitter.
Features:
Multi-Parameter, pH, ORP, Specific Ion, Dissolved Oxygen, Conductivity, Resistivity
Simple, user friendly menu structure
Noise free digital communication with sensor
Reads and writes calibration data to sensor
Dual Channel option has interactive channels, pH compensated reading, and interfering ion corrections
Non-isolated 4-20 mA output and MODBUS® RTU standard, optional HART®
Содержание DO82
Страница 1: ...HYDRA DS NO3 Rev A INSTRUCTIONAL MANUAL HYDRA DS NITRATE ANALYZER...
Страница 18: ...HYDRA DS NO3 N Page 17 Panel Mounting Cut Out 5 35 x 13 6cm x 13 6cm...
Страница 20: ...HYDRA DS NO3 N Page 19 2 4 WIRING 2 4 1 HYDRA DS TO T80 TRANSMITTER...
Страница 41: ...HYDRA DS NO3 N Page 40 Cut Out 5 35 x 5 35 13 6 cm x 13 6 cm 7 2 HYDRA DS SENSOR DIMENSIONAL DRAWING...
Страница 52: ...1500 North Kellogg Drive Anaheim CA 92807 1 714 695 0051 FAX 1 714 695 0057 ecdi com...