15

Calibration Check

Connect  a  3  litre  syringe  to  the
transducer with  the adapter supplied and
empty  by  pushing  the  handle  fully  in.
Select  ‘Calibration  Check’  from  the  main
menu and then select ‘check Calibration’.

Fill the syringe by pulling the handle at a
constant  rate  until  the  end  stop  is
reached and then immediately empty the
syringe  completely. 

Try  to  maintain  a

flow  rate  that  keeps  the  trace  within  the
grey bands on the display.

Press  ‘Reject’  to  retry  the  calibration
check at the required flow rate.

Press  ‘Again’  to  repeat  the  calibration
check at a low flow rate.

Press  ‘Again’  to  repeat  the  calibration
check at a high flow rate.

When a calibration check at all three flow
rates has been completed press ‘Done’ to
view the calibration check report screen.

Содержание CareFusion MicroLab

Страница 1: ...1 MicroLab Operating Manual English...

Страница 2: ...ting started 10 Calibration Check 15 Customisation 17 Administration Mode 18 Paper Loading 19 External printing 20 PC connection using SPCS 20 Cleaning the Transducer 22 Servicing 23 Consumables suppo...

Страница 3: ...orm of volume transducer which measures expired air directly at B T P S Body Temperature and Pressure with Saturated water vapour thus avoiding the inaccuracies of temperature corrections The transduc...

Страница 4: ...uch the name of the required region with the stylus stored on the left hand side of the unit and then touch Finish This will set the regional defaults including language if available If this screen is...

Страница 5: ...lecting from the list displayed Touch the toolbox icon to adjust volume and brightness Four levels of battery charge are indicated by the segmented battery icon When this icon turns red the battery is...

Страница 6: ...d workflow is to enter the patient s details or retrieve them from memory perform the required test and then print and save the results Please ensure that the turbine transducer is plugged in to eithe...

Страница 7: ...the help button to obtain further information To add a patient to the database use the on screen keyboard to type a unique patient ID and then touch the enter key You will then be prompted for Last Na...

Страница 8: ...H Quanjer et al Eur Respir J 1993 6 Suppl 16p5 40 Once all the patient details are added the patient is added to the database and the main menu is displayed with the patient name displayed at the bot...

Страница 9: ...iration When a manoeuvre has been obtained touch Results to view the indices Again to repeat the manoeuvre Reject to delete the manoeuvre or Done to end the test All the active indices are displayed f...

Страница 10: ...cted the default graph will be displayed This may be changed by touching the arrows at the top of the screen Flow Volume Volume time or child incentive default displays may be selected using the custo...

Страница 11: ...end stop is reached and then immediately empty the syringe completely Try to maintain a flow rate that keeps the trace within the grey bands on the display Press Reject to retry the calibration check...

Страница 12: ...ed The calibration error should be less that 3 If a greater error is shown repeat the procedure ensuring that the syringe is emptied and filled in a smooth manner without jerking the handle If an erro...

Страница 13: ...metry options allow you to configure the following Relaxed spirometry mode with or without tidal breathing Predicted value sets Predicted area or line display Display default Incentive display type Pr...

Страница 14: ...disabling of the Database Management icon will prevent the user from deleting any patient details or test results To enter administration mode turn the unit on whilst holding down the help key The def...

Страница 15: ...ce the paper into the compartment as shown and close the cover firmly It is recommended that only CareFusion thermal printer paper Cat No 36 PSA1600 is used with the MicroLab to avoid damage to the th...

Страница 16: ...shipped from the factory and should be fully charged on first use Plug the AC adapter into the mains supply and plug the adapter output plug into the power input socket on the right hand side of the...

Страница 17: ...ed to a computer that is manufactured in accordance with EN60950 1992 1993 Safety of Information Technology Equipment including Electrical Business Equipment Keep the PC out of reach of the patient at...

Страница 18: ...n below 2 Gently pull the transducer away from the housing 3 The transducer may now be immersed in warm soapy water for routine cleaning or immersed in cold disinfecting solutions e g Perasafe for a m...

Страница 19: ...y presses hold the on off button down for 10 seconds until the unit switches off and then turn on again Symbols Type B device 0086 In accordance with Directive 93 42 EEC Disposal in compliance with WE...

Страница 20: ...Filters 25 per box 36 MGF1100 MicroGard Pulmonary Filters 100 per box 36 SSC5000A PeraSafe Sterilising Powder 81g to make up 5 litres of solution 36 VOL2104 Nose Clips pack of 5 36 SM2125 3 Litre Cal...

Страница 21: ...meter Case Studies Specification of the MicroLab General Storage 2000 tests including Flow Volume loops and Volume Time curves Printer Output Compatible Hewlett Packard USB printers Specified models o...

Страница 22: ...6 FEV 75 FEV6 FEV1 as a percentage of FEV6 FEV1 FEV6 Maximum Expired Flow at 75 of FVC remaining MEF75 Maximum Expired Flow at 50 of FVC remaining MEF50 Maximum Expired Flow at 25 of FVC remaining MEF...

Страница 23: ...Relaxed vital capapcity EVC Inspiratory vital capacity IVC Breathing frequency rate FR Inspiratory time Ti Expiratory time Te Ti as a of total breath timel Ti Ttot Tidal volume as a of Ti TV Ti Tests...

Страница 24: ......

Отзывы: