From  this  menu  the  results  of  the  test
may  be  viewed,  saved,  or  printed  and
notes may be added.

It  is  also  possible  to  proceed  to  a  forced
baseline  spirometry  test,  or  a  post
medication relaxed spirometry test.

Select  ‘Exit’  when  all  the  required
functions have been used.

If  forced  spirometry  is  selected  the
default graph will be displayed.  This may
be changed by touching the arrows at the
top  of  the  screen. 

Flow/Volume,

Volume/time  or  child  incentive  default
displays  may  be  selected  using  the
customise option from the main menu.

When  the  spirometry  manoeuvre  has
been  completed  options  to  repeat  the
test, reject the test, and view results will
be available.

At  the  end  of  the  test  options  to  view
results, save results, print results, and to
add  notes  will  be  available  from  the
spirometry main menu.

Содержание CareFusion MicroLab

Страница 1: ...1 MicroLab Operating Manual English...

Страница 2: ...ting started 10 Calibration Check 15 Customisation 17 Administration Mode 18 Paper Loading 19 External printing 20 PC connection using SPCS 20 Cleaning the Transducer 22 Servicing 23 Consumables suppo...

Страница 3: ...orm of volume transducer which measures expired air directly at B T P S Body Temperature and Pressure with Saturated water vapour thus avoiding the inaccuracies of temperature corrections The transduc...

Страница 4: ...uch the name of the required region with the stylus stored on the left hand side of the unit and then touch Finish This will set the regional defaults including language if available If this screen is...

Страница 5: ...lecting from the list displayed Touch the toolbox icon to adjust volume and brightness Four levels of battery charge are indicated by the segmented battery icon When this icon turns red the battery is...

Страница 6: ...d workflow is to enter the patient s details or retrieve them from memory perform the required test and then print and save the results Please ensure that the turbine transducer is plugged in to eithe...

Страница 7: ...the help button to obtain further information To add a patient to the database use the on screen keyboard to type a unique patient ID and then touch the enter key You will then be prompted for Last Na...

Страница 8: ...H Quanjer et al Eur Respir J 1993 6 Suppl 16p5 40 Once all the patient details are added the patient is added to the database and the main menu is displayed with the patient name displayed at the bot...

Страница 9: ...iration When a manoeuvre has been obtained touch Results to view the indices Again to repeat the manoeuvre Reject to delete the manoeuvre or Done to end the test All the active indices are displayed f...

Страница 10: ...cted the default graph will be displayed This may be changed by touching the arrows at the top of the screen Flow Volume Volume time or child incentive default displays may be selected using the custo...

Страница 11: ...end stop is reached and then immediately empty the syringe completely Try to maintain a flow rate that keeps the trace within the grey bands on the display Press Reject to retry the calibration check...

Страница 12: ...ed The calibration error should be less that 3 If a greater error is shown repeat the procedure ensuring that the syringe is emptied and filled in a smooth manner without jerking the handle If an erro...

Страница 13: ...metry options allow you to configure the following Relaxed spirometry mode with or without tidal breathing Predicted value sets Predicted area or line display Display default Incentive display type Pr...

Страница 14: ...disabling of the Database Management icon will prevent the user from deleting any patient details or test results To enter administration mode turn the unit on whilst holding down the help key The def...

Страница 15: ...ce the paper into the compartment as shown and close the cover firmly It is recommended that only CareFusion thermal printer paper Cat No 36 PSA1600 is used with the MicroLab to avoid damage to the th...

Страница 16: ...shipped from the factory and should be fully charged on first use Plug the AC adapter into the mains supply and plug the adapter output plug into the power input socket on the right hand side of the...

Страница 17: ...ed to a computer that is manufactured in accordance with EN60950 1992 1993 Safety of Information Technology Equipment including Electrical Business Equipment Keep the PC out of reach of the patient at...

Страница 18: ...n below 2 Gently pull the transducer away from the housing 3 The transducer may now be immersed in warm soapy water for routine cleaning or immersed in cold disinfecting solutions e g Perasafe for a m...

Страница 19: ...y presses hold the on off button down for 10 seconds until the unit switches off and then turn on again Symbols Type B device 0086 In accordance with Directive 93 42 EEC Disposal in compliance with WE...

Страница 20: ...Filters 25 per box 36 MGF1100 MicroGard Pulmonary Filters 100 per box 36 SSC5000A PeraSafe Sterilising Powder 81g to make up 5 litres of solution 36 VOL2104 Nose Clips pack of 5 36 SM2125 3 Litre Cal...

Страница 21: ...meter Case Studies Specification of the MicroLab General Storage 2000 tests including Flow Volume loops and Volume Time curves Printer Output Compatible Hewlett Packard USB printers Specified models o...

Страница 22: ...6 FEV 75 FEV6 FEV1 as a percentage of FEV6 FEV1 FEV6 Maximum Expired Flow at 75 of FVC remaining MEF75 Maximum Expired Flow at 50 of FVC remaining MEF50 Maximum Expired Flow at 25 of FVC remaining MEF...

Страница 23: ...Relaxed vital capapcity EVC Inspiratory vital capacity IVC Breathing frequency rate FR Inspiratory time Ti Expiratory time Te Ti as a of total breath timel Ti Ttot Tidal volume as a of Ti TV Ti Tests...

Страница 24: ......

Отзывы: