XS series operating manual
XS user manual v0.55
- 103 -
0 -10 Volts analog input output (03 OPTION )
This option allows the unit to deliver a proportional analog voltage to the displayed measurements on
the LCD screen (insulation resistance, test voltages, leakage currents) and allows controlling the
dielectric strength test high voltage by an analog voltage between 0 and 10 volts. Jumpers for the
setup of the board allow selecting different features.
Specification
Output voltage
: from 0 to 10 VDC by 2.44 mV steps
Input voltage
: from 0 to 10 VDC
Output impedance
: 1 kohm ±5%
Input impedance
: 10 kohms ±5%
Insulation
: Not insulated, the ground is earthed
Connection
: screw terminal connector C6 on the rear panel
1
0-10 volts input or output
2
0-10 volts output
3
Ground
Board setup according to the unit models
The board includes some jumpers described on the schematic underneath. These jumpers allow
setup the board in a 2 analog outputs mode or in a 1 analog 1 analog input mode
Fig 12.2
SW5
SW4