17

DIAGNOSTICS cont.

To enter the keyboard diagnostic mode:

1. Turn power to the unit OFF.

Durant

®

1

Power 

In

     

2

2. Connect the rear terminal program enable input to ground.

Durant

®

Jumper 
Terminals
1 and 2

1

2

3

4

3. While holding both the right arrow key (

) and the up arrow key (

),

turn unit power ON.

Durant

Power 

In

     

The program LED and all display segments will be ON. If present, both relays will be

OFF (coils de-energized) and the analog output will be at minimum value (4 mA and

0V).

Содержание Durant Eclipse Series

Страница 1: ...umber 57700 970 04 Durant Eclipse Series Ratemeter Process Timers Models 5770X 47X Table of Contents 1 Introduction 1 Description 5 Mounting 6 Wiring 11 Programming 16 Diagnostics 19 Specifications Vi...

Страница 2: ...t per minute barrels per hour etc or in units of time such as baking time or mold ma chine cycle time In either case the Eclipse employes the fiendishly clever Tau method of calculation which means th...

Страница 3: ...e input frequency at a steady speed The Eclipse has a programmable display update time that sets the minimum time between updates Each time the display updates the unit waits until the update timer ti...

Страница 4: ...a programmable value that is common to both setpoints and both relays This means that a relay turns OFF when the display is less than or equal to the high setpoint minus the hysteresis value or when t...

Страница 5: ...al serial communication board allows a host device to download and read programming parameters and to read status information from the rate meter such as display value relay status etc This manual doe...

Страница 6: ...body until adhesive surface makes contact with the front bezel 2 Slide unit into cutout in panel 3 Attach mounting clips and screws 4 Tighten screws until unit is firmly in place DO NOT OVERTIGHTEN sc...

Страница 7: ...ree additional circuit boards with headers may be installed These option boards are relay output RS 485 serial communications and analog retransmission The option boards occupy spe cific locations in...

Страница 8: ...onnecting device for the equipment Switches and circuit breakers in Europe must comply with IEC 947 Terminal Connector Ratings AC or DC Power Input Relay Output 10A 250VAC Wire size 12 24AWG 3 1mm2 0...

Страница 9: ...ernal Fuse 1 External Fuse Size U S European 2 AMP 50 V Time Delay T2A 50 V Time Delay 12 VA AC Power Input for AC powered models 57701 47X No Internal Fuse 1 Power In L1 85 265 VAC L2 47 63 Hz 20 VA...

Страница 10: ...Max Mag Pickup Source PNP 10 kHz Max Single Ended 12 VDC Out 75 mA Max See Specifications pages 17 18 for input impedances and threshholds Relay Output Option Board Typical Wiring Durant 1 Terminal De...

Страница 11: ...on Option Board Durant 1 485 485 Common To Host Serial Port Analog Output Option Board Durant 4 20mA 1 0 10 V Common 4 20mA 0 10 V Output Ratings 4 20 mA into 750 W Ohms maximum 0 10 V into 2500 W Ohm...

Страница 12: ...to its minimum values 0V and 4 mA regardless of the input signal To enter the program mode 1 Connect Program Enable input terminal to Ground Durant Jumper Terminals 1 and 2 1 2 3 4 2 Press and hold t...

Страница 13: ...1 The parameter name is displayed by pressing and holding the program key PGM 2 The value of the parameter is displayed by releasing the program key 3 While holding the program key PGM the unit scrol...

Страница 14: ...lowing list shows all programming parameter names the default value for each and the selection range for each A programmed unit may be restored to de fault settings by pressing both the program PGM an...

Страница 15: ...scaler here The de cimal point location on this screen was selected in the scaler decimal point parameter above Range 0 001 to 9999 Display update time Sets the minimum time in seconds between displa...

Страница 16: ...log output 10V and 20 mA Range 0 to 99999 Relay 1 High Alarm Setpoint Enter the rate or time at which relay 1 will turn on if the rate or time goes above this value Range 0 to 99999 Relay 2 High Alarm...

Страница 17: ...If subsequent errors occur the NOVRAM itself may have failed and the unit should be returned to the factory for repair An error message indicates that the calibration section of NOVRAM has been corrup...

Страница 18: ...program enable input to ground Durant Jumper Terminals 1 and 2 1 2 3 4 3 While holding both the right arrow key and the up arrow key turn unit power ON Durant Power In The program LED and all display...

Страница 19: ...splay shows software part number Right Arrow All display segments and the program LED turn off and the analog output goes to maximum value 20 mA and 10V Up Arrow Each display digit will turn on one at...

Страница 20: ...ric Strength 2300 VAC to signal inputs and relays 500 VAC to RS 485 and analog outputs HUMAN INTERFACE Display 5 digits Type 56 high seven segment red LED Indicator One red LED program calibration ind...

Страница 21: ...ogrammable Parity Even odd or no parity Address Range 00 to 99 decimal Protocol Opto 22 compatible Isolation Dielectric Strength 2300 VAC to signal inputs relays and AC power inputs 500 VAC to analog...

Страница 22: ...are or may be to any extent liable including without limitation penalties imposed by the Consumer Product Safety Act P L 92 573 and liability imposed upon any person pursuant to the Magnuson Moss War...

Страница 23: ...Durant 901 S 12th Street Watertown WI 53094 800 540 9242 920 261 4070 Fax 920 261 9097 www durant com...

Отзывы: