background image

DCF-3000 Installation, Operation & Maintenance 

 

DOM0000007, Revision A 

Page 4 of 7 

 

 

Figure 2

 

 

WARNING:

 

Equipment should be shut down and 

isolated from energy sources and other equipment 
before any inspection or servicing to prevent risk of 
shock or process fluid leakage. 

 

FAILURE TO FOLLOW THIS WARNING MAY LEAD TO DEATH, 
SEVERE INJURY OR PROPERTY DAMAGE. 
 

1.

 

Disconnect all remote airlines running to the assembly, marking 
all airlines before removing. 

Isolate all sources of energy before 

disconnecting any fitting.

 Remove the center shaft connection 

bolt (Figure 1, item 12) and the four bolts (Figure 1, item 1) 
holding the twin assembly to the lid. Remove the twin assembly. 

2. 

Hand tighten packing compression wheel (Figure 1, item 6 or 
13) to  prevent lid from moving on the shaft . Remove lid 
assembly with both upper and lower cleaning discs attached

.  

3. 

Remove the 8 bolts (Figure 1, item 9) from the lid. Inspect the lid 
O-ring for damage and replace if necessary (Figure 1, item 11)

 

4. 

With the lid assembly removed, wash down the inside of vessel 
to remove all remaining process fluids and solids

5.

 

Remove ¼-20 hex nut and screw (Figure 2, item 8, 9) to 
dismantle upper and lower cleaning disc from actuator rod.  

6. 

Inspect the cleaning disc blades while inside the unit. If the 
spacing between the edge of the blade and the inside edge of 
the plate is approximately 1/8

 then the blades need to be 

replaced. See Figure 3.

                                       

7.

 

Inspect both cleaning disc (Figure 2, item 10, 11) for excessive 
wear.  If replacement is required, refer to instructions included 
with replacement disc.                                       

8.

 

After complete inspection,  and if necessary replacement, slide 
the cleaning disc on actuator rod and plug it with ¼-20 screw 
and nut. First install the upper cleaning disc (Figure 2, item 10) 
and then lower cleaning disc. (Figure 2, item 11) 

9.

 

Remove the filter element (Figure 2, item 12) from the housing 
by removing two screws (Figure 2, item 6) of element hold down 
plates (Figure 2, item 5) , and inspect the filter element for 
damage. Damage may consist of deformation, dents, holes or 
blockage. 

10.

 

Clean the element from the outside. High pressure washing may 
damage the filter element. Replace as needed 

11.

 

Remove and inspect the element O-rings (Figure 2, item 7) for 
damage and replace as needed. 

12.

 

Clean and remove any debris from the bottom of the housing. 
Inspect the purge chamber to ensure it is clean and free of large 
debris. 

13.

 

Re-install the element O-rings (Figure 2, item 7) and lower the 
element into the filter housing. When the element reaches the 
bottom, a small amount of force is needed to seat the element 
into the base. 

14.

 

Screw up the two element holding down plates to fix the element  

15.

 

Place the lid “O” ring (Figure 1, item 11) on the upper body 
flange, flat and centered.   

16.

 

Replace lid and cleaning disc assembly on the filter housing. 
Replace the lid bolt and tighten. (Figure 1, item 9) 

17.

 

Replace the twin assembly insuring that the safety device is in 
place of operating. 

18.

 

Loosen the packing compression wheel (Figure 1, item 6 or 13) 
to proper operating specification. 

19.

 

Reconnect the airlines.  

20.

 

Place the system back on line using proper plant procedures. 

21.

 

Test the safety device by tripping it while the unit is in motion. 
The unit should stop moving. If this does not work, lock the unit 
out and contact Eaton customer service. If unit operates 
correctly, reset the safety switch and put the unit back online. 

 

For maintenance of other items such as lid seals, actuators, and 
valves, see instructions supplied with repair kits. 
 

Blade removal instructions 

1.

 

Inspect the cleaning disc blades for wear and damage. 

2.

 

To replace the blades, remove the two screws holding 
the blade to the plate. Between the blade and the plate 
there is a spring. This spring must be replaced when 
the blade is replaced.  

3.

 

Place a new spring into the pocket in the new cleaning 
disc blade. Slide the blade onto the plate and insert 
the two screws into the blade. 

 

 

Figure 3 

Содержание DCF-3000

Страница 1: ...which travels across the surface of the screen These contaminants are deposited in a collection chamber at the bottom of the filter vessel where they are periodically purged out of the bottom port of...

Страница 2: ...ice is required Pressure gauges are also recommended on all process connections 2 Connect the drain line customer supplied to the filter unit purge valve To avoid restricting purge flow the drain line...

Страница 3: ...erential pressure between the inlet and outlet pressures should be used to determine stroking and purging rates Normal operation should exhibit low differential pressure of 3 5 psi that is maintained...

Страница 4: ...removing two screws Figure 2 item 6 of element hold down plates Figure 2 item 5 and inspect the filter element for damage Damage may consist of deformation dents holes or blockage 10 Clean the element...

Страница 5: ...010 250 59262 015 380 59263 024 610 59264 030 760 59265 045 1140 PACKING KITS MATERIAL STANDARD PACKING Figure 1 item 7 FLUID SEAL PACKING Figure 1 item 15 PART NO PART NO URETHANE 42958 45406 TEFLON...

Страница 6: ...must be replaced Troubleshooting Symptoms A Actuator doesn t operate properly B Purge valve will not operate properly C Leakage at lid seal D Reduction in flow rate or high differential pressure E Dro...

Страница 7: ...TIAL DAMAGES OR CONTINGENT LIABILITIES OUR MAXIMUM LIABILITY SHALL NOT IN ANY EVENT EXCEED THE CONTRACT PRICE FOR THE PRODUCT If you are interested in ordering spare parts or having service performed...

Отзывы: