background image

www.lascarelectronics.com

Issue 1   06/2022

EL-IOT-CO2

Carbon Dioxide WiFi Cloud-Connected Data Logger

Your data, anytime, anywhere

SPECIFICATIONS

CO

2

 Measurement

Range

0 to 40,000ppm

Accuracy at 400ppm

±10.5%

Resolution

1ppm

Auto-calibration

Yes

Units

ppm

Temperature Measurement

Range

-20 to +60°C (-4 to +140°F)

Accuracy

±0.3°C (±0.54°F) typical

Resolution

0.1°

Units

°C or °F

Relative Humidity Measurement

Range

0 to 100%RH

Accuracy

±2% typical

Resolution

0.1%

Dew Point Accuracy*

1.5°C typical (40 to 100%RH)

Data Logging Rate

10 seconds to 12 hours 

(user selectable)

Transmission Interval

1 minute to 24 hours 

(user selectable)

Internal Memory

300,000 readings

Dimensions

97 x 87 x 33mm

Operating Temperature

Battery Power

-18 to +55°C (0 to +131°F)

USB Power

-20 to +60°C (-4 to +140°F)

Environmental Rating

IP4X

CO

2

 FRESH AIR CALIBRATION

To maintain the accuracy of your EL-IOT-CO2, the CO

2

 sensor 

needs to recalibrate itself every 6 months. The warning icon 
will activate to remind you when it is time. Expose the device 
to fresh air by placing it outside or next to a wide open 
window. Press the CALIBRATE button to start the process, 
the display will show it is under way. After a few minutes the 
process will finish and the display will resume showing the 
current readings. If there is a problem with the calibration 
process, a warning screen will inform you. Ensure the supply 
of fresh air is adequate and try again.

*Dew point measurements are only available via the EasyLog Cloud

Содержание EL-IOT-CO2

Страница 1: ...nimum readings for carbon dioxide temperature and humidity as well as alarm and device status The EL IOT CO2 runs from a USB power supply supplied with the product and can use AA batteries to provide...

Страница 2: ...both system and data events Starter and Professional accounts are available always flexible and with no locked in contracts If you choose not to connect your EL IOT CO2 to the Cloud it will still meas...

Страница 3: ...on Not configured The logger has not yet been connected to the EasyLog Cloud readings will be displayed but not uploaded Setup Mode The device is in setup mode ready to connect to the EasyLog Cloud Co...

Страница 4: ...initiate data synchronisation with the Cloud Long Press Enter Setup mode Short Press Reset all active alarms they will immediately re trigger if an alarm threshold is breached Long Press Enter CO2 Fre...

Страница 5: ...300 000 readings Dimensions 97 x 87 x 33mm Operating Temperature Battery Power 18 to 55 C 0 to 131 F USB Power 20 to 60 C 4 to 140 F Environmental Rating IP4X CO2 FRESHAIR CALIBRATION To maintain the...

Страница 6: ...n ambient conditions the response time of the humidity sensor in your data logger is approximately 20 minutes to reach 90 of the reading However if you are measuring step changes in humidity for examp...

Страница 7: ...all Fixing Screws 3 5 x 20mm countersunk head cross recessed stainless steel BATTERYINFORMATION This device is designed to be powered using the USB power supply provided The logger does not lose its s...

Отзывы: