Environmental Requirements
• Temperature: 15
°
to 30
°
C (59
°
to 86
°
F)
Room Ambient
Temperature
• Humidity: 15% to 76%
NOTE
If the upper limit of the room ambient or the water supply temperature to
the processor is exceeded, the developer temperature may not be
controlled correctly. A water chiller may be required.
a.
Air Exhaust (full load)
Air and Heat
(1) Volume: 1.9 m
3
/min (65 ft
3
/min)
(2) Temperature: 66
°
C (150
°
F) maximum
(3) Moisture>300 gr/min or 121 gr/kg (55 gr/lb) of air
b.
Heat load to room: 4220 kJ/hr (4000 Btu/hr)
To protect the processor and equipment directly interfaced with the
processor, the dryer must be vented according to the following
specifications. Failure to properly vent the dryer exhaust can cause
corrosion within the processor and interfaced equipment. In addition, the
probability of processor-related film artifacts is increased.
[1]
The processor exhaust duct must be connected to the building exhaust ducting
system. Disposal of effluent air must comply with prevailing environmental
codes.
[2]
The following table should be used to determine the proper amount of negative
air within the duct at the end to be connected to the processor. To prevent
venturi effect at the duct opening, all measurements should be made at a point
30.5 cm (12 in.) from the open end of the duct to be attached to the processor.
Compare the average reading with the table below.
Table 4 Static Pressures
Negative Static Pressure, (Water Head)
Duct Diameter
MIN
MAX
76 mm (3 in.)
0.76 mm (0.03 in.)
1.02 mm (0.04 in.)
102 mm (4 in.)
0.25 mm (0.01 in.)
0.51 mm (0.02 in.)
1C7070
9