Electrical Requirements
IMPORTANT
Basic Requirements
In addition, all electrical services, including earth ground, must comply
with local and national electrical codes.
• 30 A, single-phase, 208/220, 3-wire, earth ground
required.
• Main Power Disconnect (wall-mounted, not furnished)
The main power disconnect switch must consist of a minimum 2-pole thermo-magnetic
circuit breaker with solid neutral and common trip, or a fused disconnect switch. This
switch must be:
• located on a wall adjacent to the processor in the lighted area
• easily accessible from the processor site
• visible from the processor site.
IMPORTANT
Standard Service
Options
In addition, all electrical services, including earth ground, must comply
with local and national electrical codes.
Table 3 Service Options
Voltage
Frequency
Service
Volts
Hz
100/200
50/60
Single-phase, 3-wire
120/208
60
Three-phase
*
, 3-wire, Wye
120/240
60
Single-phase, 3-wire
127/220
50
Three-phase
*
, 3-wire, Wye
220/380
50
Three-phase
*
, 3-wire, Wye
240/415
50
Three-phase
*
, 3-wire, Wye
220
50/60
Single-phase, 2-wire
240
50/60
Single-phase, 2-wire
*
L1, L2, and Neutral used in this configuration are sometimes referred to
as Single-Phase connections.
1C7070
7