background image

16

The front number represents the number of verification record

document, and the rear number represents the pressure point number to
be recorded. When the measurement of a certain pressure point is
completed, just long press the “

↓ “

key and the calibrator will

automatically save the current measured data, and the pressure point
number on the "verified data record identification" will increase by one
digit, waiting for the measurement of the next pressure point. When all
the pressure points to be measured are measured and recorded, the
verified data recording will be completed, and the "verified data
recording identification" on the upper part of the interface will
automatically disappear.

8.3

Verification completed

As mentioned above, when all the pressure points to be measured

are measured and recorded, the verified data recording will be
completed, and the "verified data recording identification" at the upper
part of the interface will automatically disappear. However, if the
pressure point data is not fully recorded, then the verification recording
process can be ended as follows:

Содержание ET-AY30

Страница 1: ...ET AY30 31 Pressure Calibrator User Manual Hangzhou Zhongchuang Electronics Co Ltd...

Страница 2: ...it 7 5 5 The instrument is in backlight 8 6 Menu instructions 8 6 1 Routine measurement menu 8 6 2 System setup menu 9 7 HART Function 10 7 1 HART Interface 10 7 2 HART Operation Menu 11 8 Saving the...

Страница 3: ...ransmitters with HART function and the product can also be widely used for precise measurement of pressure 1 Performance Index Pressure measuring range 100kpa 60MPa accuracy level 0 02 level 0 05 leve...

Страница 4: ...rature 30 80 Display 2 8 inch color screen 5 digit display Chinese and English can be switched Power supply built in 3 7v lithium battery power supply with 5V power adapter Auto power off function tur...

Страница 5: ...r the condition of good ventilation and no gas leakage and use the manufacturer s original power adapter Connecting RS232 communication cable in explosive gas environment is prohibited Do not replace...

Страница 6: ...the pressure measurement interface to quickly enter the pressure zero calibration interface short press the key in the menu interface to return to the previous interface ENTER enter key Long press thi...

Страница 7: ...r starting up it enters the routine measurement interface as shown in the figure the contents displayed on the screen are 1 Interface identification the interface is divided into six modes Pressure P...

Страница 8: ...ical unit of electrical logging 9 AVG MIN MAX value of pressure measurement 10 AVG MIN MAX value of electrical measurement The percentage bar of the current measured pressure relative to the full rang...

Страница 9: ...ection and select the corresponding unit directly The pressure unit will be automatically saved when you turn off the instrument and the unit displayed next time will be the unit set in the last shutd...

Страница 10: ...e figure above AVG MAX MIN record reset used to reset the AVG MAX MIN recorded value on the main measurement interface AVG MAX MIN display used to select which statistic data to display on the main me...

Страница 11: ...inutes 60 minutes Brightness control used to control the backlight brightness of the screen and there are three levels to choose from Buzzer used to select whether the buzzer will beep when the key is...

Страница 12: ...n the calibrator and the transmitter communicate successfully it will display as follows Upper limit M and lower limit M represent the maximum and minimum pressures supported by the pressure sensor in...

Страница 13: ...ured by the pressure transmitter The percentage represents the ratio of the measured pressure value to the full range The pressure value and current value under HART 2 interface represent the pressure...

Страница 14: ...ransmitter to 0 Forced output of loop current the pressure transmitter is removed from the automatic control of loop current and its output current is manually set by the calibrator with a range of 4...

Страница 15: ...lue of the calibration point it is necessary to change the set value until the actual current value is equal to the current value of the calibration point At this moment the calibration is finished Re...

Страница 16: ...menu and use the keys of to select different documents Press MODE to move the cursor to the bottom and select delete the currently selected document or create a new document The new document interfac...

Страница 17: ...the keys of to select different pressure points for checking Press MODE to move the cursor to the bottom and select edit to modify the parameters of the current document and select start to start ver...

Страница 18: ...xt pressure point When all the pressure points to be measured are measured and recorded the verified data recording will be completed and the verified data recording identification on the upper part o...

Страница 19: ...routine measurement interface will also disappear When selecting one of the pressure points in the document to view the data the display interface is shown in the figure above which can only view the...

Страница 20: ...ange an overpressure alarm will be given At this moment the pressure should be immediately reduced to avoid damage to the pressure sensor 2 If the battery life has been reduced after long use please c...

Страница 21: ...19...

Отзывы: