The material contained in this document is the property of Electronics & Innovation Ltd., it  is 
subject to change without notice.  
 
May 2016 

Revision G 

 

2.2.1 Mains Voltage 

 
The unit accommodates AC line voltages from 100 TO 240 VAC 47 

– 63 Hz 

 

2.3 OPERATION 

 
A  line  cord  is  supplied  to  form  a  connection  between  the  mains  supply  and  the 
rear  of  the  unit.  Plug  this  into  the  AC  input  at  the  rear  of  the  unit  and  the  AC 
mains outlet. 
 

2.3.1 Proceed as follows: 

 

(i) 

Ensure that there is at least 3” or 7.5 cm clearance at the rear of the 
unit for air flow. 

 

(ii) 

Ensure RF input voltage is not excessive 
a.  The 1 V rms indicated maximum input voltage is 5 times the level of 

the input signal required to achieve maximum output. Input voltages 
in excess of 2 volts peak may permanently damage the instrument. 

 

(iii) 

Connect the output via a 50 ohm coaxial lead and N plug to the load. 

 
(iv) 

Connect  the input  signal via  a 50 ohm coaxial lead and  N plug to the 
input connector. 

 

2.3.2 Front Panel Display 

 

The 3100LA front panel has a passive LCD display designed for simplicity and 
ease of use.  During initialization, the LCD shows the software revision.  After the 
amplifier is initialized, the LCD indicates Forward Power, Reflected Power, and 
amplifier status (see figure 2.1).     

Pf:___W Pr:___W 

Status:________

 

Figure 2.1 : Front Panel Display 

Where Pf refers to forward power, Pr refers to reverse power, and Status 
indicates “OK” unless there is a fault condition, such as: 

  Overheat (heat-sink temperature is too high for reliable operation) 

  PSU fault (internal fault in the main switching power supply) 

In the event of a fault, the unit may be reset by cycling the power. In the case 
over an over temp fault, ensure that the air inlet and out let are not restricted.  If 
the fault persists, please contact Service Department. 

 

Содержание 3100LA

Страница 1: ...F VOLTAGES MAY BE PRESENT AT THE OUTPUT OF THIS UNIT All operating personnel should use extreme caution in handling these voltages and be thoroughly familiar with this manual Do not attempt to operate...

Страница 2: ...ly operational such that it performs according to the original specifications free of charge to the original purchaser Should warranty service be required the unit must be returned to E I freight cost...

Страница 3: ...FM SSB pulse and any complex modulation signals The amplifier has 50 dB gain it is unconditionally stable and will not oscillate even with combinations of mismatched source and load impedance It is p...

Страница 4: ...n 3 of full scale accuracy POWER REQUIREMENTS 100 240 VAC 47 63 Hz SIZE 5 3 5 3 x 16 5 x 18 4 134 134 5 mm x 420 mm x 467 mm WEIGHT 52 pounds 23 6 kg CONNECTORS N OPERATING TEMPERATURE 0 40 C RACK MOU...

Страница 5: ...ii Connect the output via a 50 ohm coaxial lead and N plug to the load iv Connect the input signal via a 50 ohm coaxial lead and N plug to the input connector 2 3 2 Front Panel Display The 3100LA fron...

Страница 6: ...ailable terminal programs such as Hyperterm The RS 232 link has the following parameters Baud rate 19200 Data bits 8 Parity none Stop bits 1 Flow control none An example configuration using HyperTermi...

Страница 7: ...The driver output signal is fed through a length of coaxial cable to the input of the splitter network The two output signals are fed to the two power amplifiers Each amplifier has a four way splitter...

Страница 8: ...an only be externally caused by the incorrect selection of the AC supply voltage or by an input signal in excess of the specified 1 volt rms equivalent to a power level of 13dBm This chapter therefore...

Страница 9: ...bserve the gain versus frequency ripple on an oscilloscope calibrated in decibels The gain variation must be not more than 1 5 dB over the frequency range e Connect a calorimetric power meter HP434 or...

Страница 10: ...ep generator to the video sweep mode with the start frequency at 150 kHz and the sweep width to 150 MHz c Disconnect the 3100LA from the set up and connect the sweep generator RF output directly to th...

Страница 11: ...armonics Figure 4 2 1 Equipment Required a Sweep Signal Generator b Calorimetric Power Meter HP435B c Spectrum Analyzer d Attenuator 30dB e Coupler 30 dB 2 Connect the Equipment as shown in Figure 4 2...

Страница 12: ...ck Close the carton securely and seal with durable tape Mark the shipping container FRAGILE to ensure careful handling Chapter 5 Safety Do not attempt to operate this unit with the cover removed High...

Отзывы: