E8500 Instruction & Operations Manual
version 1.017
Page 43 of 49
SO
2
SENSOR
30 minutes
H
2
S SENSOR
VOC SENSOR
30 minutes
30 minutes
Span calibrate the sensor as explained in Chapter 10. If you are installing a pre-
calibrated sensor, use the following procedure:
1) While holding the
SETUP
key, press the
MEASURE
key three times. The display
will show the sensor factors.
2) Press the
DOWN
key until you reach the appropriate sensor then press OK.
3) Use the
UP / DOWN
keys to enter the correct factor, digit by digit starting with the
hundreds digit, press
OK
to move through the tens, ones, and tenths digits.