CHAPTER 3
BASIC INSTRUMENT OPERATION
It is possible to master the basic operation of the instrument in a few minutes by
following the procedure outlined below. Please refer to the other sections of this manual
for a description of the more advanced features.
The E8500 Plus emissions analyzer consists of the following three major components:
1. The probe sampling line, whose function is to extract the sample, sense
the stack temperature and depending on the option measure the stack gas
velocity
2. The conditioning system is located inside the analyzer and consists of the
thermoelectric chiller, automatic condensate draining system, and filters
3. The main section of the analyzer that houses all the gas sensors, battery
pack, and PC board
To operate the instrument, follow the steps outlined below.
1. Turn the E8500 Plus analyzer on. The instrument pump will immediately turn on
and the E INSTRUMENTS logo will appear. Press
OK
to run the autozero cycle.
2. Attach the probe and sampling line that is supplied with your analyzer. One hose
connects to the Sample In connector, and the other hose connects to the positive
(+) pressure connector. The thermocouple connector from the stack gas
sampling probe should be connected to the Temperature 1 thermocouple
connector on the analyzer.
3. If you are using the analyzer for the first time,
press the
SETUP
key to set the appropriate
parameters (i.e. fuel, units, etc.) for your
application. See Chapter 7 for an explanation of
each parameter. A table of the SETUP display is
shown here.
OCT 1 ‘15 12:45:00
Fuel: NATURL GAS
Temperature Units: F
Measure Units: PPM
Pressure Units: inWC
O2 Reference: TRUE
Pumps: OFF
Dilution Duty: 100%
Water Drain: 25min
Chiller Duty: 75%
Dilute CO: 5000 PPM
Use CO-IR: 7000 PPM
Thermal Eff: 0.30
Display Contrast: 24
Baudrate: 9.6 kbps
Version: 1.00
Battery: x.xx V