MAINTENANCE
7.1. Removal of the Sensing Probe of the Gauge Mounting Block
Gauge mounting block without hot tap valve
Make sure that the pipe line is depressurized!
•
Loose the retainer nut and pull the sensor probe from the gauge mounting block.
•
Operation without the flow sensor installed page 10.
7.2. Removal of the Sensing Probe of the Gauge Mounting Block with Hot Tap Valve
The sensor can be removed under pressure without flow interruption.
•
Turn the gauge mounting block with hot tap valve with the hand without tools (anti-clockwise rotation) in
the position SERVICE - CLOSE.
•
Only if the gauge mounting block with hot tap valve is fully in the position SERVICE - CLOSE the sensor
is hermetically sealed against the pressurized line.
•
Loose the retainer nut slowly. Let the air exhaust slowly and make sure that is was only the relief pres-
sure and the gauge mounting block with hot tap valve is hermetically sealed.
•
Loose the retainer nut and pull the sensor probe from the gauge mounting block with hot tap valve.
•
Operation without the flow sensor installed page 10.
The sensor head is sensitive. Use the protection cap for transport and storage!
7.3. Cleaning of the Sensing Element
Do not use an abrasive cleaning agent, an organic solvent containing halogen or acetone.
•
Clean the head of the sensor probe by carefully swirling in warm water of isopropyl alcohol. It is
recommended to use isopropyl alcohol if the pollution is crease or oil.
The sensor should not be touch by fingers or solid objects like screwdrivers or brushes!
•
Leave the sensor to dry in ambient air.
8.
REPLACEMENT PARTS / ACCESSORIES
8.1. Order Code Replacement Sensor
Replacement Sensor
PE772-
Type
Compact ri-le
flow direction right to left
T19
Compact le-ri
flow direction left to right
T20
Remote probe
T3
Measuring range
High - 0...200 m/s
(656.2 ft/s)
HV33
HV33
Measurement valve for
pipe diameter
DN40
N40
N40
DN50
N50
N50
DN65
N65
N65
DN80
N80
N80
Mounting
Gauge mounting block
TG2
TG2
Gauge mounting block with hot tap valve
TG3
TG3
Electrical connection
Cable gland and screw terminals
no code
1 plug for power supply and outputs
E4