User Manual EE680 

Air Velocity and Temperature Sensor for Laminar Flow

1  General

This user manual serves for ensuring proper handling and optimal functioning of the device. The user 

manual shall be read before commissioning the equipment and it shall be provided to all staff involved 

in transport, installation, operation, maintenance and repair. The user manual may not be used for the 

purposes of competition without the written consent of E+E Elektronik and may not be forwarded to 

third parties. Copies may be made for internal purposes. All information, technical data and diagrams 

included in these instructions are based on the information available at the time of writing.

Disclaimer

The manufacturer or his authorized agent can be only be held liable in case of willful or gross 

negligence. In any case, the scope of liability is limited to the corresponding amount of the order issued 

to the manufacturer. The manufacturer assumes no liability for damages incurred due to failure to 

comply with the applicable regulations, operating instructions or the specified operating conditions. 

Consequential damages are excluded from the liability.

1.1  Explanation of Symbols

This symbol indicates safety information.

It is essential that all safety information is strictly observed. Failure to comply with this information can 

lead to personal injuries or damage to property. E+E Elektronik assumes no liability if this happens.

This symbol indicates instructions.

The instructions shall be observed in order to reach optimal performance of the device.

1.2  Safety Instructions

1.2.1  General Safety Instructions

 

ƒ

The probe shall not be exposed to unnecessary mechanical stress.

 

ƒ

The sensing element

 is sensitive to electrostatic discharge (ESD), avoid touching the sensing 

element whenever possible, appropriate protective measures should be taken when touching it.

 

ƒ

Installation, electrical connection, maintenance and commissioning shall be performed by qualified 

personnel only.

 

ƒ

Use the EE680 only as intended and observe all technical specifications.

 

ƒ

Do not use EE680 in explosive atmosphere or for measurement of aggressive gases.

 

ƒ

Do not apply the nominal voltage to the RS485 data lines.

 

ƒ

This device is not appropriate for safety, emergency stop or other critical applications where device 

malfunction or failure could cause injury to human beings.

1.2.2  Intended Use

The EE680 is intended for monitoring laminar flows fulfilling highest requirements for the use in GMP 

Grade A and B or ISO class 5 clean rooms and above. It measures the standardized air velocity (vn) 

and the temperature (T) highly accurate. Its GMP-compliant design makes it the ideal solution for 

branches like microelectronics, pharmaceutical industry and other life science fields. The EE680 is 

not designed for outdoor applications. It is suitable for use in clean gases (air) free from oils, residue-

forming substances or abrasive particles.

The use of the EE680 in any other way than described in this manual bears a safety risk for people and 

the entire measurement installation and is therefore not allowed. 

The manufacturer cannot be hold responsible for damages as a result of incorrect handling, installation, 

and maintenance of the equipment.

In order to avoid damage to the instrument or health hazards, the measuring equipment must never be 

manipulated with tools that are not specifically described in this manual.

The sensor may only be utilized in accordance with the conditions defined in the technical data. 

Otherwise, measurement inaccuracies will occur and equipment failures cannot be ruled out.

Содержание EE680

Страница 1: ...BA_EE680 v1 0 Modification rights reserved User Manual EE680 Air Velocity and Temperature Sensor for Laminar Flow...

Страница 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Страница 3: ...asurement 12 5 4 3 Reference Values for the Standardized Air Velocity 12 5 4 4 Digital Interface Configuration 13 5 4 5 Selection between Analogue and Digital Output 13 5 4 6 Optical Status Indication...

Страница 4: ...ng element is sensitive to electrostatic discharge ESD avoid touching the sensing element whenever possible appropriate protective measures should be taken when touching it Installation electrical con...

Страница 5: ...or start up a device supposed to be faulty Make sure that such devices are not accidentally used by marking them clearly as faulty A faulty device may only be investigated and possibly repaired by qu...

Страница 6: ...ut or on the RS485 interface with Modbus RTU protocol The analogue output and other settings can be configured with the EE PCS 3 2 Dimensions L according to ordering code 9 7 0 4 8 0 3 89 3 5 12 0 47...

Страница 7: ...g HA010705 Connection cable HA010819 820 821 1 Supply voltage 24 V DC 20 Class III brown 2 Analogue out 2 or RS485 D Data white 3 GND blue 4 Analogue out 1 or RS485 D Clock black 5 Configuration pin g...

Страница 8: ...r For convenient handling the cable should extend 5 cm 2 in from the flange and there should be enough space for it behind the flange Connect the sensor with the M12 connector Insert the sensor into t...

Страница 9: ...y configuration by the user The EE680 factory setup corresponds to the type number ordered Please refer to the data sheet at www epluse com EE680 The user can change the factory setup with the help of...

Страница 10: ...ge from www epluse com Configurator Please refer also to chapter 5 4 4 2 Modbus protocol in the register 60001 0x00 and 60002 0x01 See Modbus Application Note AN0103 available at www epluse com EE680...

Страница 11: ...s from 0 3 Standardized air velocity vn at standard conditions factory setup Tn 23 C 73 F pn 1 013 25 hPa 14 7 psi settable via EE PCS 4 Examples For scale 100 the reading of 2550 means a value of 25...

Страница 12: ...surement of the standardized air velocity is stopped below the switch off value and resumed as soon as the switch off value plus the hysteresis is exceeded 5 4 3 Reference Values for the Standardized...

Страница 13: ...ital interface settings The interface parameters for Modbus communication can be changed Fig 13 shows the recommended values they are set in the factory 5 4 5 Selection between Analogue and Digital Ou...

Страница 14: ...Failure None No suitable supply status indication disabled GREEN Probe status normal operation no failure Laminar flow status measured value in range YELLOW FLASHING Probe status normal operation no f...

Страница 15: ...0 36 m s 70 9 ft min Range upper limit 0 54 m s 106 3 ft min Tab 4 Factory default values for laminar flow monitoring Fig 17 Response time setting The fastest response time possible is 1 5 s with ave...

Страница 16: ...oint vn T adjustment With 2 point adjustment precise measuring results can be obtained over the entire vn T measuring range The more complicated 2 point adjustment procedure is preferable to 1 point a...

Страница 17: ...17 User Manual EE680 Air Velocity and Temperature Sensor for Laminar Flow 6 2 Velocity Adjustment Fig 18 vn 1 point adjustment...

Страница 18: ...18 User Manual EE680 Air Velocity and Temperature Sensor for Laminar Flow Fig 19 vn 2 point adjustment...

Страница 19: ...19 User Manual EE680 Air Velocity and Temperature Sensor for Laminar Flow 6 3 Temperature Adjustment Fig 20 T 1 point adjustment Fig 21 T 2 point adjustment...

Страница 20: ...sensing head may be cleaned as described below and the device may be re adjusted as described in chapter 6 7 1 Cleaning The EE680 was optimized in means of used materials probe and sensor head made o...

Страница 21: ...lth of our employees at any time 7 3 Repairs Repairs may be carried out by the manufacturer only The attempt of unauthorized repair excludes any warranty claims 7 4 Error Indication on the Analogue Ou...

Страница 22: ...Accessories data sheet Modbus configuration adapter HA011018 E E Product Configuration Software EE PCS free download www epluse com configurator Protection cap M12 female connector HA010781 Protectio...

Страница 23: ...ts 1 Modbus address 68 General Power supply Class III 24 V DC 20 Current consumption typ 30 mA Electrical connection M12x1 5 poles stainless steel 1 4404 Protection rating IP65 Enclosure material Stai...

Страница 24: ...lin 69130 Ecully Tel 33 4 74 72 35 82 E mail info epluse fr E E Elektronik Germany Sch ne Aussicht 8 C 61348 Bad Homburg Tel 49 6172 13881 0 E mail info epluse de E E Elektronik Italy Via Alghero 17 1...

Отзывы: