background image

18

Hardware

Under certain circumstances, the capacitive humidity sensor element can get damaged. To avoid
the costly return of the entire transmitter to the manufacturer it is possible to replace the sensor.

Note: 

This will invalidate the factory calibration.

The sensor elements should be touched by the lead wires only. (use tweezers!)

Sensor replacement procedure:
1) Switch off supply voltage.
2) Unscrew the filter cap.
3) Pull out the humidity sensor element with a tweezer.
4) Put in the new humidity sensor - the active side (side with the sensor pads) has to

face the inside. (see diagram)

5) Screw on the filter cap. (in case of pollution replace it by a new filter cap)
6) Establish connection to PC. (RS232)
7) Switch on the supply voltage.
8) Start configuration software on PC.
9) For further instructions, see Configuration software, chapter 5.3 "Sensor/Probe replacement"

active side

Transmitters of the EE36 series are available with an optional remote sensor probe that
can be plugged into the middle section of the housing. If the sensor probe is damaged
(damage to the cable, mechanical destruction of the sensor probe) it is 
possible to replace the probe.

Note:  
This will invalidate the factory calibration.

Sensor probe replacement procedure:
1) Switch off supply voltage.
2) Remove damaged sensor probe.
3) Plug replacement probe onto middle section of the housing.
4) Establish connection to PC (RS232).
5) Switch on power supply voltage.
6) Start configuration software on PC.
7) For further instructions, see Configuration software, chapter 5.3 "Sensor/Probe replacement"

solder
points

8.

MAINTENANCE

8.3

Sensor probe replacement / optional

8.2

Sensor replacement

8.4

Fuse replacement

If the green LED on the PCB is not flashing with the supply 
voltage switched on, check the fuse and replace if required.

Fuse secondary: 

250mA / T UL248-14

Nominal voltage: 

250V 

Replacement types: 
Series: MSTU 250  

Manufacturer: Schurter

Order No.: 0034.7109

Series: 374 

Manufacturer: Littelfuse

Order No.: 374 0250

Cleaning of humidity and temperature sensors from oil residue:

1) Remove filter cap carefully, do not touch the sensors
2) Emerge the measuring head in N-HEPTAN and swirl for approx. 30 seconds 
3) Remove excess liquid and allow to air dry for approx. 30 minutes
4) Screw on filter cap carefully

Cleaning of the measuring head is recommended before emerging in other oil and 
before a calibration.

8.1

Sensor cleaning

Содержание EE36 Series

Страница 1: ...NUAL Hardware and Software Hardware and Software Series EE36 Series EE36 TRANSMITTER FOR TRANSMITTER FOR MOISTURE MOISTURE CONTENT CONTENT IN OIL IN OIL BA_EE36_06_e technical data are subject to chan...

Страница 2: ...d in accordance with the installation manual may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this...

Страница 3: ...it board 9 5 2 Display module option 10 6 ALARM MODULE OPTION 11 7 HUMIDITY TEMPERATURE CALIBRATION 12 7 1 2 point humidity calibration 12 7 2 2 point temperature calibration 13 7 3 1 point humidity c...

Страница 4: ...owed Grip sensors only at the lead wires Installation electrical connection maintenance and commissioning should be performed by qualified personnel only Safety instructions for use of the alarm modul...

Страница 5: ...s The measuring medium e g oil shall be as clean as possible i e without contamination 3 2 1 General safety instructions for installation Because the sensing probe can be exposed to very high pressure...

Страница 6: ...Installation of the probe by utilizing the ball valve set The ball valve set allows for the removal of the probe without interrupting the process Install the balve valve perpendicular to the direction...

Страница 7: ...Attention do not bend connection cable 2nd step Slowly loosen the lock nut with a spanner spanner width 24 until the expulsion force acts on the probe In the installed state never completely remove th...

Страница 8: ...Euro Standard 4 3 Connection configuration of bottom part of the housing with plug connections 8 35V DC 12 30V AC option C03 C07 The cable should be connected according to the number stamped in the pl...

Страница 9: ...ired configuration 5 OPERATING COMPONENTS I U I U Hardware OUT2 OUT1 OUT2 OUT1 1 Current voltage output 2 RS232 3 Display 4 Push button 5 Diagnosis LEDs 4 5 Connection configuration of bottom part of...

Страница 10: ...the button once again for five seconds 3 2 If both buttons are pressed for at least five seconds while the maximum value is displayed the MAX symbol disappears the maximum value will be deleted Reset...

Страница 11: ...6A Minimum load 100mA 12V Schaltpunkt switching point C kj kg g kg g m mbar Relais Status relay status 70 62 Ausschaltpunkt switching off Hysterese Einschaltpunkt switching on EIN ON AUS OFF 8 NC NC...

Страница 12: ...e performed directly on the circuit board or using the configuration software supplied for more details see Configuration Software chapter 5 4 Calibration Hardware 1 Insert the sensor probe into the r...

Страница 13: ...anged D1 flashes resp disappears when pressing alternating S1 resp S2 10 BUTTON S1 store Pressing the button for 5 seconds stores the calibration value and the procedure is ended LED D2 flashes to ind...

Страница 14: ...isappear from the optional LC display low calibration point 6 Insert the sensor probe into the reference temperature 2 high calibration point and stabilise for at least 20 minutes 7 BUTTON S1 Pressing...

Страница 15: ...5 seconds starts the procedure The calibration mode is indicated by the lit LED D2 and the symbol CALIB HIGH will appear on the optional LC display CP 50 RH or BUTTON S2 Pressing the button for 5 sec...

Страница 16: ...3 BUTTON S1 Pressing the button for 5 seconds starts the procedure The calibration mode is indicated by the symbol CALIB HIGH on the optional LC display CP 45 degC 113 F or BUTTON S2 Pressing the but...

Страница 17: ...ON S2 Pressing the button for 5 seconds starts the procedure for the calibration mode RH Pressing buttons S1 and S2 simultanously for 10 seconds customer calibration settings are reset to factory cali...

Страница 18: ...n of the sensor probe it is possible to replace the probe Note This will invalidate the factory calibration Sensor probe replacement procedure 1 Switch off supply voltage 2 Remove damaged sensor probe...

Страница 19: ...ftware Transmitter failure no supply voltage Check wiring and supply voltage only green LED is illuminated continuously Electronics defect contact the manufacturer 8 6 Replacement of sealing element B...

Страница 20: ...consumption 2x voltage output for 24V DC AC typ 40mA 2x current output typ 80mA for 100 240VAC typ 2VA Pressure range sensing pobe 0 01 10bar 0 15 145psi System requirements for software WINDOWS 98 or...

Страница 21: ...equirements MS WINDOWS 98 or higher RS232 serial interface 2 INSTALLATION Insert the CD ROM supplied with the transmitter into the PC and open the set up application Follow the instructions of the dia...

Страница 22: ...ions Save Workspace and Open Workspace apply to the tree structure only not to the configurations of individual transmitters 3 2 Interfaces Select Selects the serial interface COM port for communicati...

Страница 23: ...itter is created in the tree This procedure requires the input of a number of parameters Group Assigns a transmitter to a group Network This function is not available for the EE36 series Interface Sel...

Страница 24: ...restarts the microprocessor of the selected transmitter 3 5 Information Version Displays the version number of the EE36 software currently installed and the contact information for E E Elektronik 4 I...

Страница 25: ...caling of the output The limits must fall within the operating range indicated above Units Selects between SI or US units 5 2 Relay Used to set both optional alarm outputs Physical Quantity Selects th...

Страница 26: ...Replacement 3 Enter the nominal capacity C76 and the humidity coefficient in the corresponding input fields 4 Save the settings by clicking on the button Save Transmitter Replacement Probe 1 Open the...

Страница 27: ...after that exhaust the air around the probe 0 5h Attention Other solvents than above mentioned can corrode the humidity sensor 1 point calibration Humidity Fast and easy calibration for accurate measu...

Страница 28: ...value in the input field Humidity Reading with the reference humidity Value of the saline solution or display of HUMOR 20 8 By clicking on Save the humidity reading of the transmitter will be adjusted...

Страница 29: ...ible Type Name of the transmitter series Humidity Calibration Date Provides information on the date of the last humidity calibration but only if the configuration software was used Note Calibrations p...

Страница 30: ...tter in the Name field Complete the New Transmitter process by clicking on the button Add 6 2 How to read the configuration of a transmitter The current configuration of the selected transmitter can b...

Страница 31: ...31...

Страница 32: ...8 Bei Chen Dong St Chao Yang District Beijing 100101 P R China Tel 86 10 84992361 86 10 84992362 Fax 86 10 84992363 info epluse cn E E FRANCE Le Norly III 136 chemin du Moulin Caron F 69130 Ecully T...

Отзывы: