background image

7

Operating manual EE211 Heated humidity and temperature sensor

3.4  LED Indication

Green LED

Information during normal operation

Blue LED

 

Information during setup with the optional USB  

configuration adapter HA011066 and the EE-PCS  

Product Configuration Software

on

= everything OK

on

= USB stick connected to PC, no communication 

with EE-PCS

flashing = the main board does not recognize the  

measurement electronics inside the RH probe

flashing = communication in progress with EE-PCS

off

= no power supply or main board failure

off

= USB stick not connected to PC

Tab. 1 

LED Indication

3.5  Display

Factory Setup:

The display shows the two parameters selected for output 1 and output 2 (according to ordering code). 

For the Modbus version the display shows RH and T.

User Setup:

The user can change the display layout to 1, 2 or 3 lines and select the parameters to be displayed by 

using EE-PCS Product Configuration Software (free download from www.epluse.com/configurator) and 

the optional USB configuration adapter HA011066 (not included in the scope of supply).

3.6  Selection Output Signal Voltage / Current

The factory setup of the output signal and scaling corresponds to the type number as ordered.

The output signal voltage (U) or current 3-wire (I) can be selected with the DIP switch on the main elec-

tronics board (see 3.3 Connection Diagram - EE211-M1A2/3/5/6). This does not impact on the scaling 

of the outputs, which can be changed using the EE-PCS Product Configuration Software and the optio-

nal USB configuration adapter HA011066.

3.6.1  Examples

Factory setup: voltage output 0-5 V or 0-10 V corresponds to 0…100% RH.

After switching from U to I: current output 4 … 20 mA (3-wire) corresponds to 0 … 100% RH.

A change of the current output range for instance to 0-20 mA can be made subsequently with the EE-

PCS Product Configuration Software and the optional USB Configuration Adapter HA011066.

Factory setup: current output 0-20 mA or 4-20 mA (3-wire) corresponds to -10…50 °C.

After switching from I to U: voltage output 0-10 V corresponds to -10…50 °C.

A change of the voltage output range for instance to 0-5 V can be made subsequently with the EE-PCS 

Product Configuration Software and the optional USB configuration adapter HA011066.

4  Assembly and Installation

•  Insert the M20 cable gland included in the scope of supply into the corresponding opening of the 

EE211 basis unit and fix it tight with the nut. Alternatively, it is possible to install a 1/2” conduit fitting 

(not included in the scope of supply).

•  Install the EE07-MT temperature probe either directly onto the M12 connector of the EE211 basis unit 

or using the optional probe cable.

The entire EE211 shall be places inside the space to be monitored. The measured RH correspond to 

the location of the T probe. Consequently, the T probe shall be positioned at the place of main interest 

for RH measurement. For instance, in an environmental chamber, the EE211 basic unit can be fixed 

conveniently on the inside wall, while the T probe can be placed in the middle of the chamber using the 

optional probe cable.

Содержание EE211

Страница 1: ...BA_EE211_e V1 3 technical data are subject to change Operation Manual EE211 HEATED HUMIDITY TEMPERATURE SENSOR...

Страница 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Страница 3: ...7 3 6 1 Examples 7 4 Assembly and Installation 7 5 Setup 8 5 1 EE PCS Product Configuration Software 8 5 2 Digital Interface RS485 with Modbus RTU Protocol 8 5 2 1 Hardware Bus Termination 8 5 2 2 Add...

Страница 4: ...ed due to failure to comply with the applicable regulations operating instructions or the operating conditions Consequential damages are excluded from the liability 1 1 Explanation of Symbols This sym...

Страница 5: ...hem clearly as faulty A faulty device may only be investigated and possibly repaired by qualified trained and authorized staff If the fault cannot be fixed the device shall be removed from the system...

Страница 6: ...ensing head E E proprietary coating of the sensing element and sealed solder pads minimize the impact of corrosive agents Maximum humidity of 76 RH at the sensor eliminates the drift caused by exposur...

Страница 7: ...using the EE PCS Product Configuration Software and the optio nal USB configuration adapter HA011066 3 6 1 Examples Factory setup voltage output 0 5 V or 0 10 V corresponds to 0 100 RH After switching...

Страница 8: ...the USB configura tion adapter HA011066 PC HA011066 Fig 3 Configuration adapter 5 1 EE PCS Product Configuration Software 1 Download the EE PCS Product Configuration Software from www epluse com confi...

Страница 9: ...set via 1 EE PCS Product Configuration Software and the appropriate configuration cable 2 Modbus protocol in the register 60001 0x00 and 60002 0x01 See Application Note Modbus AN0103 available on www...

Страница 10: ...ter number1 Dec Register address2 Hex Parameter name 60001 0x00 Slave ID5 modbus address 60002 0x01 Modbus protocol settings6 Tab 5 Integer 16 bit read and write register INFO read register Function c...

Страница 11: ...dress Hi Starting address Lo No of register Hi No of register Lo CRC Request Hex EF 03 00 19 00 02 EB 7B Response Hex EF 03 04 38 5F 41 D5 0A E3 Modbus ID address Function code Byte count Register 1 v...

Страница 12: ...roves the measurement accuracy of a device The specimen is compared with the reference and brought in line with it An adjustment can be followed by a calibration which documents the accuracy of the ad...

Страница 13: ...e a Windows PC using the EE PCA product configuration adapter and the cable HA011057 see data sheet EE PCA EE PCS free download from www epluse com configurator shall be installed on the PC Start EE P...

Страница 14: ...lsoure material Polycarbonate UL94V 0 with Display UL94HB approved T probe material Stainless steel 1 4571 Protection class IP65 NEMA 4 Cable gland M20 x 1 5 Electromagnetic compatibility EN61326 1 In...

Страница 15: ...15 Operating manual EE211 Heated humidity and temperature sensor...

Страница 16: ...0 6460 Office Stuttgart Tel 49 151 538 37 500 E E Elektronik Italy info epluse it Tel 39 02 2707 86 36 E E Elektronik France info epluse fr Tel 33 4 74 72 35 82 E E Elektronik USA office epluse com Of...

Отзывы: