13
Operating manual EE211 Heated humidity and temperature sensor
6.5 Individual Calibration / Adjustment of the T Probe
The T probe can be also adjusted separately, which might be desirable in high demanding applications.
For best accuracy it is advisable to use a dry block calibrator.
6.5.1 Adjustment Procedure
• Connect the T probe to the a Windows PC using the EE-PCA product configuration adapter and the
cable HA011057 (see data sheet EE-PCA).
• EE-PCS (free download from www.epluse.com/configurator) shall be installed on the PC.
• Start EE-PCS and proceed with the calibration or adjustment of the T reading.
Please note:
In case of individual adjustment, the adjustment data is stored in the T probe, the factory setup of the
EE211 basis unit is not altered. Consequently, the T probe can be readjusted or replaced at any time
without the need of readjusting the EE211 basis unit.