5.0 Smartblower Functional Description
By pressing the power button/menu button the blower is switched on. By pressing the power button for 2 seconds,
the blower is switched off again.
Upon switching on, an acoustic signal sounds briefly and the display lights up.
5.1 Switch-on Phase/System Test
Afterwards the blower performs a self-check. In this phase, the start logo
is displayed with the selected SVE mode (a). This process takes about 5
seconds.
If an error occurs during the Self-Check phase, an alarm is shown on the
display with the corresponding error code for 5 seconds. The unit then
switches itself off automatically to prevent the blower from starting up
without correcting the error.
If no error occurs during the Self-Check phase, the display automatically
switches to the main menu.
(a)
Symbols
Meaning
One Filter Mode:
• SVE mode is set to particle filter in one-filter mode.
Can optionally be operated with Y-connector and two particle filters.
• SVE mode is set to head parts for hood mode.
•
Compatible Head Parts:
PM Lab Hood AV, PM Chemical Hood and e-breathe Multi-Hood
•
Compatible Filter
: Can only be operated with particle filters.
Two Filter Mode:
• SVE mode is set to combination filter in two filter mode.
• SVE mode is set to head parts for hood mode.
•
Compatible Head Parts
:
PM Labhood AV*, PM Chemical Hood* and e-breathe Multimask**
•
Compatible Filter
:
*Can only be used with two combination filters.
**Can only be used with two particle filters and can not be used with combination
filters.
Full Face Mask Mode:
• SVE mode is set to particle filter in one-filter mode.
• SVE mode is set to head parts for full mask mode.
•
Compatible Head Parts
: PM Full Face Mask Panarea
•
Compatible Filter
: Can only be operated with particle filters.
5.1.1 Start Logo / SVE-Mode Display
The default SVE mode is displayed at the start logo. Depending on the application, this mode is predefined by the
manufacturer for the head section and the filter used. Only the manufacturer or a authorized service partner can
change the SVE mode using a PC application.
Depending on the head piece used, a differentiation is made between hood mode and full mask mode for the SVE
modes.
In hood mode, a distinction is made between one-filter and two-filter operation. The corresponding mode depends
on the number and selection of filters.
The full mask mode is only possible with one filter mode, since this mode is only intended for use with a particle
filter.
Einfilter-Betrieb
Einfilter-Betrieb
Zweifilter-Betrieb
Vollmasken-
Betrieb