persists, contact DYSIS Medical
service department.
Weekly calibration is
required
Calibrate the system.
Image is displayed, but the lights do not turn
on
Illumination source
malfunction
Contact DYSIS Medical service
department.
The DYSIS Ultra 2.0 Digital Colposcope
doesn’t
turn on
Main power cable is
disconnected
Ensure the power switch is OFF
and reconnect the main power
cable to the power brick and the
power cable is locked into the
DYSIS Ultra 2.0 Digital
Colposcope base. Turn the main
power switch ON.
Outlet has no electricity
Ensure the electrical outlet is
operational.
System malfunction
Contact DYSIS Medical service
department.
Calibration cannot be performed
The card is very dirty
Contact DYSIS Medical service
department.
System malfunction
Contact DYSIS Medical service
department.
Software update cannot be performed
The USB stick is not
recognized
Use another USB stick formatted
to FAT32.
The configured network
path with the upgrader
is not accessible
Check connections through the
connectivity tab in SETTINGS to
ensure the configured network
path is accessible.
The DYSIS Ultra 2.0
Digital Colposcope
software upgrade is not
on the USB stick
Follow software update
instructions carefully. The
upgrade file should not be
embedded in a folder on the
USB.
System malfunction
Contact DYSIS Medical service
department.
Power cable cannot be plugged in to DYSIS
Ultra 2.0 Digital Colposcope
The inner circle is
pressed, and power
cable cannot lock into
DYSIS Ultra 2.0 Digital
Colposcope
Hold the outer circle and pull
back the inner circle to connect
the power cable.