3.)
After the LOAD HI prompt, AUTO LOWER appears. If set to YES, the lift
system will lower the weight onto the piston after the melt time expires. If set to
NO, the lift must be manually told to lower the weight via the keypad.
4.)
After the AUTO LOWER prompt, AUTO RAISE appears. If set to YES, the
lift system will raise the weight after the run is complete. If set to NO, the lift
must be manually told to raise the weight via the keypad.
5.)
Proceed with run normally.
Series 4000 Melt Indexer Operation Manual
“Flow Ratio” Runs
40