COMM PORT
MINI LIFT
PNEU LIFT
LOCKOUT
DATE and TIME
MACHINE ID
LANGUAGE
ORIFICE DIA
ORIFICE LEN
Introduction
Create (New)
Recall (Old)
Save Program
While EDITING
[ PC, OFF, DIAG] Open up communications port for computer output, no
output or diagnostics mode, respectively.
[OFF or ON] Enable the mini lift to function during material runs when
installed.
[OFF or ON] Enable pneumatic lift to function during material runs, when
installed.
[OFF or ON] Enable password protection. When ON users can not edit test or
system parameters only run tests. The default password is 1234, if the
LOCKOUT is enabled this number must be put in to turn the LOCKOUT off.
The NUMBER item allows a new password number to be entered.
Set the appropriate date and time here. Follow prompts.
[##] Assign a unique machine number to the instrument. If using MIWORKS
be sure this is set to one if you have only a single machine or to the Porter box
port number for multiple machines. This allows you to tell what machine your
sample was run on when you have more than one machine.
[CUSTOM, ENGLISH, FRENCH, ITALIAN, SPANISH, GERMAN] Switch
between desired languages.
[XXX] Store standard orifice diameter, 2.096 mm. Used in calculations.
[X.XXX] Store standard orifice radius, 8.000 mm. Used in calculations Scroll
through this menu and make appropriate changes. Press ESC twice and any
changes that have been made will be implemented.
Programming
This section is intended for a quick reference of method parameters. Please refer
to the appropriate sections in the manual for more specific information about a
particular method.
Before a measurement can be made, the RUN PARAMETERS must be
entered. These parameters dictate how the experiment will be performed. Many
of these parameters are set by particular ASTM test procedures. The SERIES
4000 software has many of these values pre-loaded into memory.
Press the
EDIT
SMART Key to start the EDIT mode. Enter the new program
number into the upper left hand corner of the display and press ENTER.
Proceed to EDIT.
Press the PROG button and enter the program number in the upper left hand
corner of the display or BROWSE with the Smart Keys. Press ACTIVE to use
the selected program number.
From the EDIT mode, ESC out until prompted to SAVE PROGRAM. Select
SAVE to save.
In order to receive the SAVE PROGRAM prompt, a value
must be changed while in the editing mode
.
Browse the suggested values appearing on the 3rd line of the Edit Screen with
the
SMART Key (
Browse Key
). Select the desired value with the
Series 4000 Melt Indexer Operation Manual