Selecting TEST in the SYS Mode
TEST:RTD Display
TEST:Digital Encoder
TEST:Printer
TEST:Display Brightness
Introduction
REJECTION
AVERAGE
MFR / MVR
B FLAG
VISCOSI
TY
PRINTE
R
FORM FEED
Press SELECT to view actual temperature reading accurate to 1/100 of a degree
Celsius. Dynisco Polymer Test technical support may have you do this under
certain conditions. RTD Display Test may also come in handy when
equilibrating RTD’s and thermometers.
Displays the current digital encoder arm position. Readings for home position
should be 105.00 mm, +/- 1.5 mm.
Your model must support printer output to perform this test. A test pattern
output will be produced on the connected printer. Press STOP to end printer
output.
The operator can control the VFD brightness level. Select a percentage of
default brightness.
The SETUP mode
This software routine is used to set user preferences for data collection. This
routine also turns off instrument options such as the digital encoder or pneumatic
lift. Please scroll through the settings prior to commissioning the instrument and
make a note of the values particular to your instrument. If for some reason the
settings are changed or erased they can be easily re-entered from this menu. By
pressing the SETUP key the following series of inputs are requested of the user.
Press the ENTER SMART Key to input the appropriate value. After entry is
complete, press the ESC key to save modified information.
After pressing SETUP, scroll through the list of settings using the UP or
DOWN SMART KEYS . Select the EDIT SMART KEY to change the parameter
accordingly. Please reference
Configuring the Instrument Section
for a
detailed description of each parameter. Note the order the items appear can vary.
[1..5, OFF] ] Exclude data points that are more than the operator entered
number of standard deviations from the mean. Exclude these points from all
statistical calculations. Points are not deleted but marked on the printouts with
asterisks. The rejection option is used for systems configured with a printer.
Data points are printed but not included in statistics.
[AUTO, OFF, ON]
ON:
average is computed for one or more data collection
runs. Generates statistics for printout.
AUTO:
compute the average of a single
data collection run without operator interaction at the end of the run.
[MFR or MVR] The operator can choose the calculated result. The choices are
Melt Flow Index or Melt Volume Index respectively.
[ENCODR or FLAG] Enable either the Digital Encoder or the Optical Flag
devices, respectively.
[POISE, PA-S, REYN] Select units of viscosity for displayed results.
[MINI, OFF, EPSON, IBM, LASER] Select output device. Mini printer, No
device, Epson compatible, HP Laser Jet compatible, respectively. (Only 4002
models will have the mini-printer option)
[OFF or ON] Perform a Form Feed (A page is sent out of the printer) after the
run data is printed.
Series 4000 Melt Indexer Operation Manual
Configuring the Instrument
17