Operational Problems
Symptom
Probable Cause Resolution
Aspiration is not
complete.
(continued)
Hydrophobic filter
may be
obstructed.
Replace filter.
High Waste Level
is displayed.
The Waste
Container is full.
Empty the Waste
Container (see page 70).
The waste level
sensor cable is
not plugged in.
Connect the waste level
sensor cable (see
page 17).
The waste level
sensor connector
is defective.
Service is required.
The waste level
sensor detector is
reversed.
Service is required.
The waste
container is
deformed.
The waste tubing
is blocked.
Remove the waste tubing
from the rear of the
instrument (see page 17).
Check the waste tubing
(and aspiration tubing)
and remove any
blockages or kinks.
The waste
container is
damaged or
defective.
Replace the waste
container.
X Motor Home
Error
is displayed.
Motion of the
wash head is
being blocked.
Check the wash head
vicinity and remove any
obstacles.
The motor is
defective.
Service is required.
Z Motor Home
Error
is displayed.
Motion of the
wash head is
being blocked.
Check the wash head
vicinity and remove any
obstacles.
The motor is
defective.
Service is required.
Test Too Large
is
displayed.
The wash
protocol exceeds
the memory limit.
Reduce the number of
steps in the wash protocol
and try again.
80
Opsys
MW™
User’s Guide
Содержание Opsys MW
Страница 1: ...Opsys MW Microplate Washer User s Guide IMPORTANT Please read carefully before using the Opsys MW ...
Страница 4: ......
Страница 8: ...About this Manual This page is intentionally left blank iv Opsys MW User s Guide ...
Страница 14: ...Description This page is intentionally left blank 6 Opsys MW User s Guide ...
Страница 16: ...Features This page is intentionally left blank 8 Opsys MW User s Guide ...
Страница 20: ...Warning Labels This page is intentionally left blank 12 Opsys MW User s Guide ...
Страница 46: ...Configuring the System This page is intentionally left blank 38 Opsys MW User s Guide ...
Страница 74: ...Washing a Plate This page is intentionally left blank 66 Opsys MW User s Guide ...
Страница 84: ...Pump Calibration This page is intentionally left blank 76 Opsys MW User s Guide ...