Configuring the System
Specifying the Pump Parameters
The CAL. command allows the calibration values for each pump to be
specified:
1. Select SETUP (if needed). The display will read:
Select option
HEAD
PLATE
2. Scroll right to display the next selection:
Select option
VERSION CAL.
3. Select the pump pressure (high or low). The display will read:
Pump speed? 190
4. Enter the desired value using the keypad and press ENTER. The
display will read:
Slope? 2.1
Note
: The Slope and Intercept values should
only be changed by
DYNEX Technologies
Service Personnel.
5. Press ENTER to keep the existing value. The display will read:
Intercept? 24
6. Press ENTER to keep the existing value. The display will again
read:
Select option
VERSION CAL.
Note
: Refer to page 74 for additional pump
calibration instructions.
36
Opsys
MW™
User’s Guide
Содержание Opsys MW
Страница 1: ...Opsys MW Microplate Washer User s Guide IMPORTANT Please read carefully before using the Opsys MW ...
Страница 4: ......
Страница 8: ...About this Manual This page is intentionally left blank iv Opsys MW User s Guide ...
Страница 14: ...Description This page is intentionally left blank 6 Opsys MW User s Guide ...
Страница 16: ...Features This page is intentionally left blank 8 Opsys MW User s Guide ...
Страница 20: ...Warning Labels This page is intentionally left blank 12 Opsys MW User s Guide ...
Страница 46: ...Configuring the System This page is intentionally left blank 38 Opsys MW User s Guide ...
Страница 74: ...Washing a Plate This page is intentionally left blank 66 Opsys MW User s Guide ...
Страница 84: ...Pump Calibration This page is intentionally left blank 76 Opsys MW User s Guide ...