Installing the Waste Container
Installing the Waste Container
To install the Waste Container and Foam/Aerosol Trap:
1. Insert the Quick-Connect end of the vacuum tubing on the
Foam/Aerosol Trap to the Quick-Connect vacuum connector at
the rear of the instrument Figure 3).
2. Screw on the Waste Container cap.
3. Route the waste tubing through the lower slot at the front corner
of the instrument and through the routing clips at the front of the
instrument.
4. Connect the end of the waste tubing to the waste tubing
connector on the wash head.
5. Connect the waste level sensor cable to the waste level sensor
connector at the rear of the instrument (Figure 3).