Option Card
Calibration Menu
(Pg 2 of 2)
Enter the option card calibration menu from the calibrate configuration menu.
This menu is used to calibrate the analog inputs for current and voltage control
and the analog outputs for current and voltage readings.
The upper voltage control analog input is calibrated at 100% of
the analog control signal range (20mA for a 4-20mA signal
range). A control signal set to 100% of the specified range
should be connected. When the RMT button is pressed this
calibration value is automatically updated. Refer to the
calibration procedure for details on calibrating.
The upper voltage reading analog output is calibrated at 100%
of the analog control signal range (20mA for a 4-20mA signal
range). When the RMT button is pressed this calibration value
is adjusted until the analog output is equal to 100% of the analog
control signal range. Refer to the calibration procedure for
details on calibrating.
The lower voltage control analog input is calibrated at 10% of
the analog control signal range (5.6mA for a 4-20mA signal
range). A control signal set to 10% of the specified range
should be connected. When the RMT button is pressed this
calibration value is automatically updated. Refer to the
calibration procedure for details on calibrating.
The lower voltage reading analog output is calibrated at 10% of
the analog control signal range (5.6mA for a 4-20mA signal
range). When the RMT button is pressed this calibration value
is adjusted until the analog output is equal to 10% of the analog
control signal range. Refer to the calibration procedure for
details on calibrating.