Page 60 of 126
198-0853-02 Rev F
6.
After completing a calibration the calibration values should be saved by selecting “yES”
when prompted to store changes. The calibration values are stored to permanent memory
and will be retained through power on and power off cycles.
Cal
Point
Calibration Menu Display
Calibration point(s) Description
“LCL” Calibration (Read back meters
and output)
CL01
“I.Hi Out CL01”
Upper current output
CL02
“I.Hi mtr CL02”
Upper current read back meter
CL03
“I.Lo Out CL03”
Lower current output
CL04
“I.Lo mtr CL04”
Lower current read back meter
CL05
“E.Hi Out CL05”
Upper voltage output
CL06
“E.Hi mtr CL06”
Upper voltage read back meter
CL07
“E.Lo Out CL07”
Lower voltage output
CL08
“E.Lo mtr CL08”
Lower voltage read back meter
“OPt” Calibration (Analog Option
Card)
CL09
“I.Hi CtL CL09”
Upper analog current control input (optional)
CL10
“I.Hi rEAd CL10”
Upper analog current reading output
(optional)
CL11
“I.Lo CtL CL11”
Lower analog current control input (optional)
CL12
“I.Lo rEAd CL12”
Lower analog current reading output
(optional)
CL13
“E.Hi CtL CL13”
Upper analog voltage control input (optional)
CL14
“E.Hi rEAd CL14”
Upper analog voltage reading output
(optional)
CL15
“E.Lo CtL CL15”
Lower analog voltage control input (optional)
CL16
“E.Lo rEAd CL16”
Lower analog voltage reading output
(optional)