No. 60027, Issue 4. January 1998
7 Programming
45
7.5
Calibrating the SLM using the HHP
There are twelve SLM parameters that must be calibrated to the individual
wheelchair system using the HHP. These parameters are all described in section
7.3.
The first four parameters calibrated by the HHP are approached sequentially. The
eight Restrictor Plate parameters are grouped and calibrated in one operation of
the HHP.
7.5.1
Initial Operation
1. Plug HHP into the Programmer Socket on the DX Remote and turn the DX
System on. The initial screen appears for two seconds.
Dynamic DX
Programmer
VERSION 1.10
2. Then the main menu screen reads :
* * MAIN MENU * *
View or edit?
Program : 1 ?
NEXT
YES
3. Press 'NEXT' until the Technician Mode screen appears :
* * MAIN MENU * *
Technician Mode
disabled. Enable ?
NEXT
YES