No. 60027, Issue 4. January 1998
7 Programming
41
Stall Time
The stall timeout delay. If the current limit is
exceeded for more than this time a Stall Timeout
Range 5 - 50 seconds
Fault (SLM Flash Code 11) will be displayed by
Accessed by : DCL
the SLM (refer to Diagnostics section 8).
Stall Timeout
If set to ‘yes’, driving is disabled after Stall Time if
in current limit.
State yes / no
Accessed by : DCL
A Stall Timeout Fault (Flash Code 11) is displayed
by the SLM (refer to Diagnostics section 8).
I²T
If set to ‘on’, motor I²T is enabled.
State on / off
I²T parameters define the thermal characteristics of
Accessed by : DCL / OEM
the motors used in wheelchair.
The OEM must provide sample motors or detailed
motor specifications to Dynamic Controls Ltd.
Dynamic will provide recommended setting to
ensure correct motor protection against
overheating.
I²T Threshold
I²T parameter, see above.
Range 10 - 90 %
Accessed by : DCL / OEM
Motor Time Scale
I²T parameter, see above.
Range 10 - 200
Accessed by : DCL / OEM
Max Motor Temp
The maximum motor temperature used for I²T
protection of the motor.
Range 70 - 200
E
C
Accessed by : DCL / OEM
I²T parameter, see above.