The Brk/Bridge Off Delay setting will depend on how quickly the Park
Brakes engage. The Park Brakes must engage before the motors turn off. It is
safer to err on the high side.
Emergency Decel.
Emergency Decel sets the rate at which the wheelchair stops under most fault
conditions or if the user turns the DX System off while driving. If the PM
detects a fault which it considers unsafe and is unable to control, it will
automatically bring the wheelchair to a halt at this rate. The optimum value
will depend on the wheelchair type, the manufacturer’s preferences, and any
regulations that apply to the country of use.
A higher value will produce a faster emergency deceleration than a lower
value.
Warning :
Front Wheel Drive wheelchairs should have this parameter set
lower e.g. 50%. Otherwise the wheelchair will stop too quickly, and the user
may be thrown from the wheelchair.