CHAPTER 5 PARAMETER SETTING & CALIBRATION
parameter setting and calibration will be operated through wireless indicator
1.Key function
KEY
FUNCTION
【
TARE
】
Move digit
【
ZERO
】
Set value up
【
MODE
】
Storage and exit
【
ENTER
】
Confirm
2. Parameter setting
OPERATION
DISPLAY
ILLUSTRATION
Press
【
MODE
】
key
twice
〖
SETUP
〗
Confirm, enter into parameter setting program
Press
【
ENTER
】
〖
FS=06
〗
New dynamometer, it display FS=06
,
if
dynamometer calibrated, it will display capacity.
Press
【
ZERO
】
to
choose
〖
FS=05
〗
Choose capacity, the screen display02/03/05/06/10/
15/20/30/50/60/75/80 circularly
,
take cap:5000kg/2kg
as example, choose FS=05, See following list about
FS, Id, Pt chosen.
Press
【
ENTER
】
to
confirm
〖
Id=02
〗
New dynamometer , the screen display Id=02, if
dynamometer calibrated, it will display calibrated
division.
Press
【
ZERO
】
to
choose
〖
Id=02
〗
Choose division, it display 01/02/05/10/20circularly.
5000kg dynamometer division is 2kg, so choose
Id=02
Press
【
ENTER
】
〖
Pt=0
〗
New dynamometer, it display Pt=0, if dynamometer
calibrated, it will display calibrated decimal position
Press
【
ZERO
】
to
choose
〖
Pt=0
〗
Choose decimal position, the screen
display 0/1/2/3 circularly, 0=xxxxx, 1=xxxx.x,
2=xxx.xx, 3=xx.xxx
,
Cap.5000kg ,we choose Pt=0
Press
【
ENTER
】
to confirm
〖
Ab=24
〗
Display zero range , A: zero range by hand; B: zero
range automatically, 0~5 total 6 options:
0=0%F. S; 1=2%F.S; 2=4%F.S; 3=10%F.S;
4=20%F.S; 5=50%F.S,
Generally Don’t change
this parameter.
Press
【
ENTER
】
to confirm
〖
Cd=11
〗
New dynamometer, the screen display Cd=11,if
Dynamometer calibrated, it display calibrated
value.
USER MANUAL 11