16
Dymax ECE Series UV Flood-Lamp Systems User Guide
Step 1.
Turn the Power Switch off and disconnect the Power Supply from the electrical power source.
Allow the entire assembly to cool for 15 minutes.
CAUTION!
The bulb operates at temperatures exceeding 500°C. Touching the Bulb before
sufficient cool down time is allowed will cause severe burns. Always wear safety eye wear while
replacing the bulb.
Always use a soft, clean rag, paper towel, or gloves when handling the bulb. Skin oils left on the
bulb will burn into the quartz, reducing output intensity. If the bulb is inadvertently touched, clean
the Bulb thoroughly with a soft, clean rag and IPA.
Step 2.
Hold the Lamp/Reflector Assembly Housing securely and remove the Mounting Hardware from
the Lamp/Reflector Assembly Housing Bracket. Lift the housing off the
ZIP Shutter
and place it
upside down on a clean work surface.
Step 3.
Reach into the Reflector and grasp the flat area of Bulb at either end (Figure 15). Lightly push the
Bulb toward the Socket on the opposite end of the lamp so that the end being grasped can be
lifted clear of the Socket (Figure 16). Install the new Bulb by following Steps 4 & 5 in the reverse
order.
IMPORTANT:
Install the Bulb so the seal dimple on the bulge of the glass is facing towards
the Reflector surface. Avoid touching the Quartz Tube with your fingers.
NOTE:
Bulb has no polarity.
Figure 15. Flood Lamp Bulb Installed
Figure 16. Push Bulb Towards Socket
Socket
Socket
Lamp (Bulb)