BlueWave® MX-MIM User Guide
43
Table 9 Continued. Alarms Chart for BlueWave® MX-MIM
Alarm # Alarm/Warning Description
Cause/Corrective Action
103
Emitter Bad LED
Return emitter to Dymax.
104
Emitter Fan Alarm
Return emitter to Dymax.
110-113 QX4 Wand1
–
Wand 4
Temperature >90°C
Reduce intensity setting.
Improve airflow to BlueWave® QX4® wand.
Reduce ambient temperature.
114-117 QX4 Wand1
–
Wand4 Fault
Return wand to Dymax.
118-121 QX4 Wand1
–
Wand4 Lost Comms
Verify wand cable installation.
Replace wand extension cable.
Check cable connections on both wand and
controller ends.
Inspect wand connector for damage.