Dymax BlueWave® LED VisiCure® User Guide
NOTE: Multi-pole lightguides should be checked for leg-to-leg output uniformity. Rotate the
lightguide to verify the desired intensity of each leg before tightening the set screw.
■
The Lightguide is now installed and ready for use. The end of the Lightguide should be periodically
cleaned with isopropyl alcohol. Adhesive build-up may be removed from the end of the Lightguide by
gently scraping using a razor blade or razor knife.
4.
Turn the Power Switch to the on position.
5.
There is no need for a warm-up or cool-down period with this LED system. The unit may be powered
down and powered up again immediately without causing any negative effects.
6.
Activate an exposure cycle by pressing on the Footswitch. With the Timer/Manual Selector Switch
(located on the front panel of the unit) in the manual position, the exposure time is controlled directly
from the Footswitch. In the timed position, operation is determined by the setting displayed on the
electronic timer. Simply push the timer setting buttons to enter the desired number of seconds the
LED needs to be powered.
7.
With the system exposure active, adjust the Intensity Adjustment Screw as required to achieve the
desired output intensity on an appropriate radiometer. The Intensity Adjustment Screw is a 10- turn
potentiometer.
Figure 6. Intensity Adjustment Screw