Hysteresis Function: 

Keeps switching state of output stable if flow rate varies about preset
value. With flow rate rising, output switches when switch-on point has
been reached (SPx). With flow rate falling, output does not switch back
until switch-off point (rPx) has been reached. To adjust hysteresis: set
switch-on point, then set switch-off point at requested distance.

Window Function: 

Enables monitoring of a defined acceptable range. When flow rate varies
between switch-on point (SPx) and switch-off point (rPx), output is
switched (window function/NO) or not switched (window function/NC).
Width of window can be set by means of the difference between SPx and
rPx. SPx=upper value, rPx=lower value.

Scaling measuring range (analog output): 

Analog start point (ASP) is the measured value at which the output sig-
nal 4 mA is defined. Analog end point (AEP) is the measured value at
which the output signal 20 mA is defined. Minimum distance between
ASP and AEP = 25% of final measuring range value. Output signal
ranges from 4 to 20 mA in set measuring range. Output is also indicated
by: flow rate above measuring range (output signal > 20 mA), or flow rate
below measuring range (output signal between 3.6 and 4 mA).

Fault Indication

OL: Detection range exceeded (flow rate > 120% of final value of mea-
suring range)

SC 1: Flashing - short circuit in switching output 1*
SC 2: Flashing - short circuit in switching output 2*
SC: Flashing - short circuit in both switching output*
Err: Flashing - fault in measuring probe

*Concerned output is switched off as long as short circuit continues.
Faults are indicated even if display is deactivated.

Figure 7

Figure 5

Figure 6

Chart 3

C-32:Magnesense bulletin  9/24/09  11:48 AM  Page 6

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание CAM Series

Страница 1: ...Steel 304S15 ceramics glass passivat ed PEEK polyester Fluoroelastomer anodized aluminum Response Time 0 1 seconds Temperature Limit 32 to 140 F 0 to 60 C Accuracy 1 41 3 0 3 FV 3 44 6 MV 0 6 FV Pressure Limit 232 psig 16 bar Humidity Limit 90 RH Power Requirements 19 30 DC Output Signal Output 1 PNP Open Collector VOUT max 30 VDC ISINK max 250mA Output 2 4 20mA scalable or PNP Open Collector Loop...

Страница 2: ...ng the unit on its side with the pipe length to the right This may jeopardize the accuracy Mounting To achieve the specified measurement accuracy the follow ing mounting conditions must be adhered to defined inflow outflow pipe lengths defined flow cross sectional area fixed installation depth and correct positioning of the measuring elements Integrate the pipe length so that the direction of flow...

Страница 3: ...ey Functions Figure 4 Chart 2 To decrease the value cycle past the maximum value and the setting will restart from the minimum www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Страница 4: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Страница 5: ...able as the current count of the total izer possible data loss can be maximum 10 minutes Display indicates current measured values yellow LED s signal switching status of out puts Display unit can be temporarily changed press Set button briefly after 15 seconds unit returns to display unit set in menu point Uni Display Mode Indicates parameters and set parameter values When Mode Enter button is pu...

Страница 6: ...ween SPx and rPx SPx upper value rPx lower value Scaling measuring range analog output Analog start point ASP is the measured value at which the output sig nal 4 mA is defined Analog end point AEP is the measured value at which the output signal 20 mA is defined Minimum distance between ASP and AEP 25 of final measuring range value Output signal ranges from 4 to 20 mA in set measuring range Output...

Отзывы: