background image

4  DRYSTIK 

4.1  Passive DRYSTIK 

A passive DRYSTIK may be installed in the Water 
Probe system without modifying any other part of the 
system or the operating conditions. 

Th

e inner 

membrane tube goes between the exchanger and the 
desiccant while the outer sheath is purged by dry air 
from the RAD7 outlet. 

Th

e two 

ows should be in 

opposite directions along the DRYSTIK. A 12” 
DRYSTIK will increase the life of the desiccant by a 
factor of about 

ve. 

4.2  DRYSTIK ADS-3 and ADS-3R 

DURRIDGE’s DRYSTIK ADS-3 and ADS-3R models 
include a pump upstream of the inner membrane 
tube, and a needle valve downstream of the tube. 

Th

is 

increases the pressure inside the tube, which 
increases its e

ciency. 

A typical setup has the RAD7 pump set to OFF 
(

Setup, Pump, Off [ENTER]

), the DRYSTIK 

pump running continuously and the needle valve 
adjusted to give a 

ow rate of about 0.2 L/min. 

When using an active DRYSTIK, a Small Drying 
Tube or larger Laboratory Drying Unit full of 
desiccant can be added to keep the air sample in the 
RAD7 below 7% RH. It should be inserted between 
the DRYSTIK air outlet and the inlet 

lter on the 

RAD7, as shown in Fig. 4 on the following page. 

Th

desiccant will last for a very long time, and the drying 
tube will add only a tiny volume to the air loop. 

4.3  E

ect on Response Time 

With a 

ow rate of only 0.2 L/min it will take about 

20 minutes for the air in the loop to go around once. 

Th

is will add an extra 10 or 15 minutes to the 

response time for radon. For long term studies the 
slower response is generally not important, whereas 
the frequency of replacing the desiccant may be. So 
an active DRYSTIK may be of considerable bene

t. 

4.4  Custom DRYSTIK Settings 

Th

e standard 0.2 L/min 

ow rate of the DURRIDGE  

DRYSTIK is typically used because it matches the 
average 

ow rate of a RAD7 in AUTO mode, and it 

also matches the performance of the installed pump 
at a pressure of 44 PSI (3 atmospheres). DURRIDGE’s 
Active DRYSTIK models can be set to maintain a 44 
PSI pressure inside the inner membrane tubing and a 

ow rate of 1L/min or even more, to restore the speed 

of response of the system while virtually eliminating 
the need to periodically replace the desiccant. See the 
DRYSTIK user’s manual for details. 

Section 4 

 

DRYSTIK

11

Содержание RAD7

Страница 1: ...WATER PROBE Passive Slow Response Radon in Water Accessory for the RAD7 User Manual...

Страница 2: ...Protocol 6 1 4 3 User Protocol 6 2 MEASUREMENT PROCEDURE 7 2 1 Start Up 7 2 1 1 Temperature Probe 7 2 1 2 Start Measurement 7 2 2 Speed of Response 7 2 2 1 Measurement in Progress 7 2 2 2 In fl uenci...

Страница 3: ...ff ect on Response Time 11 4 4 Custom DRYSTIK Settings 11 Fig 4 Water Probe con fi guration with Active DRYSTIK 12 5 CARE MAINTENANCE AND TROUBLESHOOTING 13 5 1 Water Catastrophe 13 5 2 RAD7 Care 13...

Страница 4: ...mperature is measured with a temperature probe inserted into the water As an example at a water temperature of 10 degrees Celsius the partition coe ffi cient is about 1 3 Th at means there is three ti...

Страница 5: ...t this end to the air inlet fi lter with 1 8 ID tubing at the fi lter end which is then placed on the RAD7 INLET Th e Luer taper ensures an airtight connection Fig 2 RAD7 Water Probe Standard Setup 1...

Страница 6: ...tteries in a fully charged condition Set the RAD7 s Mode to Auto which starts the instrument in Sni ff mode counting 218 polonium decays only and switched automatically to Normal Mode counting 218 pol...

Страница 7: ...a standard deviation of 10 At very low concentrations it may take hours and averaging over many cycles to reach a su ffi ciently precise value 2 2 2 In fl uencing Factors Th ere are two processes req...

Страница 8: ...the process and which must be replenished Th e most obvious is the desiccant A new or regenerated Laboratory Drying Unit will normally last for about ten days of continuous use in a temperate climate...

Страница 9: ...ce as instructed Th e water temperature information is used by CAPTURE to calculate the radon in water concentrations For detailed instructions please see the Water Probe information in the CAPTURE us...

Страница 10: ...formula is applied automatically in CAPTURE when the Water Type is set to Fresh Water in the Run Parameters Window as shown in Fig 3 below 3 2 2 Schubert Et Al Formula If the Water Type is set to Sal...

Страница 11: ...It should be inserted between the DRYSTIK air outlet and the inlet fi lter on the RAD7 as shown in Fig 4 on the following page Th e desiccant will last for a very long time and the drying tube will ad...

Страница 12: ...Fig 4 Water Probe con fi guration with Active DRYSTIK Section 4 DRYSTIK 12...

Страница 13: ...id onto its hinges Th e wrap will make the RAD7 almost watertight If it tears it can be easily replaced at any time Th e instrument should in any case be returned every year for recalibration It is us...

Страница 14: ...dioxide measurements in coastal waters Environmental Science and Technology 46 14 7685 7691 Schubert M Paschke A Lieberman E Burnett W C 2012 Air Water Partitioning of 222Rn and its Dependence on Wat...

Страница 15: ...978 667 9556 Fax 978 667 9557 Web www durridge com Email service durridge com Copyright 2021 DURRIDGE Company Inc All rights reserved DURRIDGE the DURRIDGE logo and the Works with the DURRIDGE logo ar...

Отзывы: