background image

2  MEASUREMENT PROCEDURE 

2.1  Start Up 

2.1.1  Temperature Probe 

If you are using a Temperature Probe (sold 
separately), load the temperature logger so

ft

ware and 

connect the Temperature Logger to the PC using the 
serial cable provided. Con

gure the logger to take 

temperature readings at frequent intervals (these may 
be far more frequent that the RAD7 test cycles.)  

Choose the second (external) temperature sensor. 
Connect the Temperature Probe to the logger and 
note that when you hold the probe the indicated 
temperature rises. 

When everything is set, begin recording. 

Th

e logger’s 

LED may 

ash periodically. Once the logger has 

begun running, you may remove the serial cord from 
both the logger and the PC. 

Warning!

  

Make sure that previous temperature data 

has been downloaded before launching the logger. 

Th

e launching process may erase previous data. 

Insert the Temperature Probe into the water next to 
the Water Probe. 

2.1.2  Start Measurement 

Switch on the RAD7 (have the printer switched on if 
you are using it. 

Th

e RAD7 will then print a header 

for the data printout, including a review of the setup, 
before it gives you a ‘Test’ prompt.) 

Provided that the RAD7 has been set up properly (see 
above), at the ‘

Test

’ prompt, push 

[ENTER]

 then the 

right arrow, to see ‘

Test Start

’ on the LCD, then 

push 

[ENTER]

 to start the test. 

2.2  Speed of Response 

2.2.1  Measurement in Progress 

Th

e instrument is now measuring the radon in the 

closed air loop that includes the Water Probe’s 
membrane tubing. With high radon activity 
concentration in the surrounding water it will take 
half an hour or more before there is much of a 
reading, and 2 - 3 hours before you can rely on the 
count rate being close to the equilibrium value. A

ft

er 

that you need to accumulate su

cient counts for the 

precision desired. For example, 100 counts would 
give a reading with a standard deviation of 10%. At 
very low concentrations, it may take hours, and 
averaging over many cycles, to reach a su

ciently 

precise value. 

2.2.2  In

uencing Factors 

Th

ere are two processes requiring time. One is for the 

air in the closed loop to approach equilibrium with 
the surrounding water and the other is for the RAD7 
to respond to the changed radon concentration in the 
air loop. 

Th

rst is a property of the exchange tubing 

and the second is determined by the half life of the 

rst daughter of radon, namely 218-polonium (Sni

 

mode), or the half lives of intermediary 214-lead and 
214-bismuth (Normal mode). See the RAD7 manual, 
Section 3 for more details. 

2.2.3  Air Flow Rate 

Th

ere is also a small contribution to the response 

time from mixing of the radon-laden air around the 
closed air loop. 

Th

is depends on the duty cycle of the 

RAD7’s pump. However, even with the pump set to 
AUTO, this delay is only around 20 minutes, which is 
small compared with the 2-hour timescale for radon 
exchange across the membrane tubing and the 3-hour 
95% response time of the RAD7 running in Normal 
mode. For this reason, it is almost always su

cient to 

set the pump to AUTO mode, in which the air is 
circulated around the closed air loop for one minute 
in every 

ve. Air will remain stationary in the Water 

Probe for four minutes before moving to the 
Laboratory Drying Unit where it waits another four 
minutes before entering the RAD7. While in the 
Water Probe, the stationary air may approach 
equilibrium with the water thus inhibiting further 
radon transfer from the water to the air. It will be 
about 15 to 20 minutes before that parcel of air 
returns to the Water Probe. We can therefore estimate 
that the response time of the system will be increased 
by about 20 minutes if the pump is set to AUTO 

Having the pump set to AUTO would normally be 
associated with having the RAD7 cycle time of 30 
minutes or longer. So an extra 15 to 20 minutes on 
the response time will not be excessive. 

Section 2 

 

Measurement Procedure

7

Содержание RAD7

Страница 1: ...WATER PROBE Passive Slow Response Radon in Water Accessory for the RAD7 User Manual...

Страница 2: ...Protocol 6 1 4 3 User Protocol 6 2 MEASUREMENT PROCEDURE 7 2 1 Start Up 7 2 1 1 Temperature Probe 7 2 1 2 Start Measurement 7 2 2 Speed of Response 7 2 2 1 Measurement in Progress 7 2 2 2 In fl uenci...

Страница 3: ...ff ect on Response Time 11 4 4 Custom DRYSTIK Settings 11 Fig 4 Water Probe con fi guration with Active DRYSTIK 12 5 CARE MAINTENANCE AND TROUBLESHOOTING 13 5 1 Water Catastrophe 13 5 2 RAD7 Care 13...

Страница 4: ...mperature is measured with a temperature probe inserted into the water As an example at a water temperature of 10 degrees Celsius the partition coe ffi cient is about 1 3 Th at means there is three ti...

Страница 5: ...t this end to the air inlet fi lter with 1 8 ID tubing at the fi lter end which is then placed on the RAD7 INLET Th e Luer taper ensures an airtight connection Fig 2 RAD7 Water Probe Standard Setup 1...

Страница 6: ...tteries in a fully charged condition Set the RAD7 s Mode to Auto which starts the instrument in Sni ff mode counting 218 polonium decays only and switched automatically to Normal Mode counting 218 pol...

Страница 7: ...a standard deviation of 10 At very low concentrations it may take hours and averaging over many cycles to reach a su ffi ciently precise value 2 2 2 In fl uencing Factors Th ere are two processes req...

Страница 8: ...the process and which must be replenished Th e most obvious is the desiccant A new or regenerated Laboratory Drying Unit will normally last for about ten days of continuous use in a temperate climate...

Страница 9: ...ce as instructed Th e water temperature information is used by CAPTURE to calculate the radon in water concentrations For detailed instructions please see the Water Probe information in the CAPTURE us...

Страница 10: ...formula is applied automatically in CAPTURE when the Water Type is set to Fresh Water in the Run Parameters Window as shown in Fig 3 below 3 2 2 Schubert Et Al Formula If the Water Type is set to Sal...

Страница 11: ...It should be inserted between the DRYSTIK air outlet and the inlet fi lter on the RAD7 as shown in Fig 4 on the following page Th e desiccant will last for a very long time and the drying tube will ad...

Страница 12: ...Fig 4 Water Probe con fi guration with Active DRYSTIK Section 4 DRYSTIK 12...

Страница 13: ...id onto its hinges Th e wrap will make the RAD7 almost watertight If it tears it can be easily replaced at any time Th e instrument should in any case be returned every year for recalibration It is us...

Страница 14: ...dioxide measurements in coastal waters Environmental Science and Technology 46 14 7685 7691 Schubert M Paschke A Lieberman E Burnett W C 2012 Air Water Partitioning of 222Rn and its Dependence on Wat...

Страница 15: ...978 667 9556 Fax 978 667 9557 Web www durridge com Email service durridge com Copyright 2021 DURRIDGE Company Inc All rights reserved DURRIDGE the DURRIDGE logo and the Works with the DURRIDGE logo ar...

Отзывы: