If the Natural Rock Sample has been calibrated by DURRIDGE Company, the calibration certi

cate 

gives the standardized average of readings 3, 4, 5 and 6, obtained with the standard setup, a well purged 
and dry RAD7 and 30-minute cycles. 

Th

us the ratio of the value stated on the rock sample calibration 

certi

cate to the measured standardized average gives you the correction factor directly. RAD7 

readings multiplied by that ratio will give readings consistent with the DURRIDGE calibration system 
to within 15%. 

You may also, if you wish, calculate the total radon amount in the air loop, and hence originally in the 
Natural Rock Sample container. If the radon amount is Rn and the volume of the air loop is V, then the 
radon concentration is Rn/V. 

Th

e air volume in the Natural Rock Sample is about 460ml, a standard 

RAD7 is 800ml, and the tubing with small drying tube is 16ml, for a total volume of a standard Natural 
Rock Sample system with standard RAD7 of 1.275 L. If Rn is expressed in pCi, then the radon 
concentration will be Rn/1.275 pCi/L. If Rn is expressed in Bq, then the radon concentration will be 
Rn/1.275 Bq/L, or 1000 * Rn / 1.275 Bq/m

3

In pCi, Rn = 1.275 * radon concentration in pCi/L. 

In Bq, Rn = 0.001275 * radon concentration in Bq/m

3

 

If the certi

ed and observed standardized average values di

er by more than 15%, we recommend that 

the RAD7 be returned to DURRIDGE Company for service. If, a

ft

er several monthly sensitivity checks, 

there is a consistent ratio between the observed and the certi

ed standardized average radon 

concentration, you may consider applying a multiplier to all readings made with that RAD7, to bring 
them in line with the Natural Rock Sample standard. If you notice a big jump, up or down, in the 
reading, from the previous month, 

rst look to see if the RAD7 was properly purged before the 

measurement was started. If there is no obvious reason for the jump, the RAD7 should be returned to 
DURRIDGE for service. 

Figure 4  Graph of Natural Rock Sample sensitivity test data 

Sec$on 2

 

 

RAD7 Radon Sensitivity Check 

 

 

 

 

 

 

      

7

Содержание NATURAL ROCK SAMPLE

Страница 1: ...NATURAL ROCK SAMPLE Performance Monitoring Accessory for the RAD7 User Manual...

Страница 2: ...2 3 1 Drying and Resealing the Natural Rock Sample 5 2 4 Radon Data Assessment 6 2 5 Multiple RAD7s 8 2 6 Cumulative Spectrum 9 3 RAD7 THORON SENSITIVITY CHECK 10 3 1 Thoron Measurement 10 3 1 1 Radon...

Страница 3: ...the closed loop radon calibration circuit or upstream of the desiccant in the open loop thoron calibration con guration If the desiccant in the Small Drying Tube gets used up it should be replaced or...

Страница 4: ...be compared with the known concentration and thus the sensitivity of the RAD7 may be e ectively checked A correction may be applied to the Natural Rock Sample radon reading to account for the ambient...

Страница 5: ...Setup Protocol Recycle 06 ENTER Put the RAD7 s infrared printer in place and switch o the RAD7 Switch on the RAD7 printer Switch on the RAD7 and let it print out the header including a review of the s...

Страница 6: ...re one month from that date The next measurement should be made within 50 days of the date on the label to avoid the risk of moisture creeping in and changing the emanation e ciency and hence the rado...

Страница 7: ...ock Sample system with standard RAD7 of 1 275 L If Rn is expressed in pCi then the radon concentration will be Rn 1 275 pCi L If Rn is expressed in Bq then the radon concentration will be Rn 1 275 Bq...

Страница 8: ...al Rock Sample and the radon distributed around the loop the radon concentration will be 2 1 5 1 333 Bq L 1333 Bq m3 If every additional RAD7 adds 1L to the loop then four RAD7s would total 4 5L inclu...

Страница 9: ...d with all pumps except one set to OFF 7 Depending on the precision required after between 3 and 24 hours the tests may be stopped and the RAD7s disconnected 8 If required rough thoron checks may then...

Страница 10: ...measurement The setup assumed in the RAD7 data processing and stated as standard in the manual consists of a Small Drying Tube which may be used as a wand for sni ng along with a standard input tubing...

Страница 11: ...Sample it will make no di erence to the average radon reading but will make a big di erence to the thoron readings Even just switching the connections to the Natural Rock Sample can a ect the thoron r...

Страница 12: ...Figure 7 Thoron Sensitivity Check con guration Sec on 3 RAD7 Thoron Sensitivity Check 12...

Страница 13: ...n check the rate of injection of thoron into the sample ow will always be the same Let this be Th The units of Th are pCi per minute or Bq per minute If the RAD7 ow rate measured at the RAD7 outlet is...

Страница 14: ...978 667 9556 Fax 1 978 667 9557 Web www durridge com Email service durridge com Copyright 2020 DURRIDGE Company Inc All rights reserved DURRIDGE the DURRIDGE logo and the Works with the DURRIDGE logo...

Отзывы: