background image

2.1  Purge the RAD7

 

First, purge the RAD7 with dry air in an open loop for at least 5 minutes, then switch to 1-day protocol: 
select Setup, Protocol, 1-day, and push [ENTER]. Start a reading, still sampling fresh air through a 
drying unit. Go to the third status window (push the [Menu] key, then push [ENTER], [ENTER], Rt 
Arrow, Rt Arrow, and observe the relative humidity in the top right corner. Keep the reading going 
until the RH is below 9%, then switch o

 the RAD7. 

If a correction is to be made to the Natural Rock Sample radon reading for radon added to the air loop 
by the ambient air, a

ft

er purging the RAD7, (see section 2.4), the fresh-air measurement should be kept 

going until at least one and preferably two or three cycles are completed. 

2.2  Hook up to the Natural Rock Sample

 

Hook the RAD7 up to the Natural Rock Sample in a closed loop, using the tubing and Small Drying 
Tube provided with the Natural Rock Sample. Use the con

guration shown in Figure 2, with the Small 

Drying Tube upstream of the RAD7, and the Natural Rock Sample connection upstream of the Small 
Drying Tube. If one end of the Small Drying Tube has started to turn pink, indicating that it is wet, that 
end should be positioned closest to the RAD7 outlet. 

 
2.3  Start a 3-Hour Test

 

Switch on the RAD7 and press the menu key. Select 1-day protocol (

Setup,

 

Protocol,

 

1-day, 

[ENTER]

). Change the Recycle number to 6 (

Setup,

 

Protocol,

 

Recycle,

 

06,

 

[ENTER]

). Put the 

RAD7’s infrared printer in place and switch o

 the RAD7.

 

Switch on the RAD7 printer. Switch on the RAD7 and let it print out the header, including a review of 
the settings. Check the printed review to make sure the Cycle time is 00:30 (30 minutes) and the 
Recycle number is 06. Open the ball valves on the Natural Rock Sample container and start a test 
(

Test, Start,

 

[ENTER]

). 

Keep the measurement going until all six cycles are complete and the summary is printed out (3 hours). 
Average the last four readings. Record that average with the date. If a thoron test is to be conducted (see 
Section 3), the best time to do it would be 1 - 2 hours a

ft

er the radon check (see Section 3.1.1 for 

details). 

2.3.1  Drying and Resealing the Natural Rock Sample 

A

ft

er testing is complete, the Natural Rock Sample should be thoroughly dried and prepared for 

storage. To do this, set up the Natural Rock Sample in a closed loop with a RAD7 and a laboratory 
drying unit containing fresh desiccant. (At this point the Small Drying Tube that was used in the 3-
hour test should be sealed with yellow caps and set aside until the next 3-hour test is conducted.) 

Th

Natural Rock Sample should be positioned downstream of the laboratory drying unit, and the 
laboratory drying unit should be connected with the end furthest from the screw cap connected to the 
Natural Rock Sample. 

Th

e screw cap end, which may be pink (indicating that it is wet), should be 

connected to the RAD7 outlet as shown in Figure 3, on the following page. 

Sec$on 2

 

 

RAD7 Radon Sensitivity Check 

 

 

 

 

 

 

      

5

Содержание NATURAL ROCK SAMPLE

Страница 1: ...NATURAL ROCK SAMPLE Performance Monitoring Accessory for the RAD7 User Manual...

Страница 2: ...2 3 1 Drying and Resealing the Natural Rock Sample 5 2 4 Radon Data Assessment 6 2 5 Multiple RAD7s 8 2 6 Cumulative Spectrum 9 3 RAD7 THORON SENSITIVITY CHECK 10 3 1 Thoron Measurement 10 3 1 1 Radon...

Страница 3: ...the closed loop radon calibration circuit or upstream of the desiccant in the open loop thoron calibration con guration If the desiccant in the Small Drying Tube gets used up it should be replaced or...

Страница 4: ...be compared with the known concentration and thus the sensitivity of the RAD7 may be e ectively checked A correction may be applied to the Natural Rock Sample radon reading to account for the ambient...

Страница 5: ...Setup Protocol Recycle 06 ENTER Put the RAD7 s infrared printer in place and switch o the RAD7 Switch on the RAD7 printer Switch on the RAD7 and let it print out the header including a review of the s...

Страница 6: ...re one month from that date The next measurement should be made within 50 days of the date on the label to avoid the risk of moisture creeping in and changing the emanation e ciency and hence the rado...

Страница 7: ...ock Sample system with standard RAD7 of 1 275 L If Rn is expressed in pCi then the radon concentration will be Rn 1 275 pCi L If Rn is expressed in Bq then the radon concentration will be Rn 1 275 Bq...

Страница 8: ...al Rock Sample and the radon distributed around the loop the radon concentration will be 2 1 5 1 333 Bq L 1333 Bq m3 If every additional RAD7 adds 1L to the loop then four RAD7s would total 4 5L inclu...

Страница 9: ...d with all pumps except one set to OFF 7 Depending on the precision required after between 3 and 24 hours the tests may be stopped and the RAD7s disconnected 8 If required rough thoron checks may then...

Страница 10: ...measurement The setup assumed in the RAD7 data processing and stated as standard in the manual consists of a Small Drying Tube which may be used as a wand for sni ng along with a standard input tubing...

Страница 11: ...Sample it will make no di erence to the average radon reading but will make a big di erence to the thoron readings Even just switching the connections to the Natural Rock Sample can a ect the thoron r...

Страница 12: ...Figure 7 Thoron Sensitivity Check con guration Sec on 3 RAD7 Thoron Sensitivity Check 12...

Страница 13: ...n check the rate of injection of thoron into the sample ow will always be the same Let this be Th The units of Th are pCi per minute or Bq per minute If the RAD7 ow rate measured at the RAD7 outlet is...

Страница 14: ...978 667 9556 Fax 1 978 667 9557 Web www durridge com Email service durridge com Copyright 2020 DURRIDGE Company Inc All rights reserved DURRIDGE the DURRIDGE logo and the Works with the DURRIDGE logo...

Отзывы: