1  INTRODUCTION 

 

Th

e Natural Rock Sample is a 21oz (600g) sample of granite gravel, enclosed in a container that may be 

sealed using a pair of ball valves and end caps. 

Granite generally contains trace amounts of both uranium and thorium that has been in the rock since 
it was formed, millions if not billions of years ago. 

Th

erefore all the progeny of both the uranium and 

thorium decay chains are fully supported and in full equilibrium down to 226-Ra and 224-Ra. 

When the Natural Rock Sample is kept dry, it constitutes a rock-steady emitter of both radon and 
thoron. 

Th

is manual shows how to take advantage of that property to check the performance of a 

RAD7. Please note that to keep the emission constant it is essential that the Natural Rock Sample be 
kept fully dry. 

Th

is is because the radon and thoron emanation e

ciency depends upon the humidity at 

the rock-air interface on the surface of the rocks. Always use a Small Drying Tube, with blue, active 
desiccant in it, in the closed-loop radon calibration circuit, or upstream of the desiccant in the open-
loop thoron calibration con

guration. If the desiccant in the Small Drying Tube gets used up it should 

be replaced or regenerated. 

Th

e Natural Rock Sample should be thoroughly dried by purging with dry 

air every 30 - 50 days. 

Th

is can be done as part of the standard radon calibration check protocol (see 

Section 2.3.1) or, if no calibration check is required, as a standalone procedure. 

Th

e Natural Rock 

Sample has a desiccant pack and humidity indicator built into it for maintaining and verifying dryness. 
For thoron sensitivity measurement, the carrier air, pushing the thoron into the sample path, should be 
dried before it enters the rock sample. 

Th

ese con

gurations are illustrated on the following pages. 

A

ft

er calibration at DURRIDGE Company, and with the optional 

Th

oron Calibration Check Kit, the 

Natural Rock Sample can be used for low-precision checking of both the radon and thoron sensitivities 
of any RAD7. 

Th

e Natural Rock Sample can provide a means for monitoring long-term changes in both 

radon and thoron sensitivity. 

 

 

Figure 1  Natural Rock Sample Components 

Sec$on 1

 

 

Introduction

   

 

 

 

 

 

 

 

 

      3

Содержание NATURAL ROCK SAMPLE

Страница 1: ...NATURAL ROCK SAMPLE Performance Monitoring Accessory for the RAD7 User Manual...

Страница 2: ...2 3 1 Drying and Resealing the Natural Rock Sample 5 2 4 Radon Data Assessment 6 2 5 Multiple RAD7s 8 2 6 Cumulative Spectrum 9 3 RAD7 THORON SENSITIVITY CHECK 10 3 1 Thoron Measurement 10 3 1 1 Radon...

Страница 3: ...the closed loop radon calibration circuit or upstream of the desiccant in the open loop thoron calibration con guration If the desiccant in the Small Drying Tube gets used up it should be replaced or...

Страница 4: ...be compared with the known concentration and thus the sensitivity of the RAD7 may be e ectively checked A correction may be applied to the Natural Rock Sample radon reading to account for the ambient...

Страница 5: ...Setup Protocol Recycle 06 ENTER Put the RAD7 s infrared printer in place and switch o the RAD7 Switch on the RAD7 printer Switch on the RAD7 and let it print out the header including a review of the s...

Страница 6: ...re one month from that date The next measurement should be made within 50 days of the date on the label to avoid the risk of moisture creeping in and changing the emanation e ciency and hence the rado...

Страница 7: ...ock Sample system with standard RAD7 of 1 275 L If Rn is expressed in pCi then the radon concentration will be Rn 1 275 pCi L If Rn is expressed in Bq then the radon concentration will be Rn 1 275 Bq...

Страница 8: ...al Rock Sample and the radon distributed around the loop the radon concentration will be 2 1 5 1 333 Bq L 1333 Bq m3 If every additional RAD7 adds 1L to the loop then four RAD7s would total 4 5L inclu...

Страница 9: ...d with all pumps except one set to OFF 7 Depending on the precision required after between 3 and 24 hours the tests may be stopped and the RAD7s disconnected 8 If required rough thoron checks may then...

Страница 10: ...measurement The setup assumed in the RAD7 data processing and stated as standard in the manual consists of a Small Drying Tube which may be used as a wand for sni ng along with a standard input tubing...

Страница 11: ...Sample it will make no di erence to the average radon reading but will make a big di erence to the thoron readings Even just switching the connections to the Natural Rock Sample can a ect the thoron r...

Страница 12: ...Figure 7 Thoron Sensitivity Check con guration Sec on 3 RAD7 Thoron Sensitivity Check 12...

Страница 13: ...n check the rate of injection of thoron into the sample ow will always be the same Let this be Th The units of Th are pCi per minute or Bq per minute If the RAD7 ow rate measured at the RAD7 outlet is...

Страница 14: ...978 667 9556 Fax 1 978 667 9557 Web www durridge com Email service durridge com Copyright 2020 DURRIDGE Company Inc All rights reserved DURRIDGE the DURRIDGE logo and the Works with the DURRIDGE logo...

Отзывы: