2170100001L29 2109V005
49
Troubleshooting
EN
14.4 Error messages on the touch screen.
Problem
Probable cause
Solution
Error code 1008
Connection interrupted
i
Update the firmware
Error code
1010
Temperature of unit too high
i
Allow the unit to cool down
i
Contact technician
Error code
1022
Subassembly not initialised
i
Fault in software, update the software
if required
i
Contact technician
Error code
1024
Internal data communication
fault
i
Switch the unit off and back on again
i
Update the firmware
i
Darken the room
i
Turn the unit so that no light can fall
directly into the insertion slot
Error code
1026
Incorrect acquisition mode
i
Select a different acquisition mode
i
Inform a service technician
i
Update the firmware
i
Reset the scanning modes to the fac-
tory settings via the unit interface or
the Imaging Software
Error code
1100
Permitted time for scan pro-
cess exceeded
i
Contact technician
i
Check the belt drive
i
Check for blockage, remove image
plate from unit
Error code
1104
Erasure unit fault
i
Contact technician
i
Replace the erasure unit
Error code
1153
Unit fault
i
Switch the unit off and on again
i
Update firmware
Error code
1154
Internal data communication
fault
i
Switch the unit off and on again
i
Update firmware
Error code 1160
Final pentaprism assembly rota-
tion speed not reached
i
Contact technician
i
Update firmware
i
Replace the pentaprism assembly if the
problem occurs regularly
Error code 1170
SOL sensor timeout Fault on the
laser, SOL sensor or pentaprism
assembly
i
Contact technician
i
Update firmware
Error code 1172
SOL sensor timeout Fault on the
laser, SOL sensor or pentaprism
assembly
i
Inform a service technician
i
Update the firmware
Содержание VistaScan Ultra View