Assembly
- "Update" of devices that are connected to
the IT network
- "Upgrade" of devices that are connected to
the IT network
– The data connection utilises part of the band-
width of the network Interactions with other
medical devices cannot be completely ruled
out Apply the IEC 80001-1 standard for risk
assessment
– The device is not suitable for direct connec-
tion to the public Internet
Take care when connecting units together or to
parts of other systems as there is always an ele-
ment of risk (eg due to leakage currents)
i
Only connect units when there can be no
question of danger to operator or to patient
i
Only connect units when it is safe to do so
and when there is no risk of damage or harm
to the surroundings
i
If it is not 100% clear from the unit data sheet
that such connections can be safely made or if
you are in any doubt, always get a suitably
qualified person (eg the manufacturer) to verify
that the setup is safe
i
Observe the specifications of IEC 60601-1
(EN 60601-1) when connecting the appliance
with other appliances, eg a PC system, both
in and outside the patient environment
Only connect peripheral units (such as com-
puters, monitors or printers) that conform at
least to the requirements set out in
IEC 60950-1 (EN 60950-1)
i
The connected computer must conform to
EN 55032 (class B) and EN 55024
Connecting the unit via the network cable
i
Remove the cover from the rear of the device
i
Connect the supplied network cable to the
network connection (item 2) of the device
2
3
1
The cover on the rear must be correctly
fitted when the device is operated within
the patient environment
EN