Electromagnetic compatibility (EMC)
Interference immunity measurements on the supply input
Immunity to conducted disturbances, induced by radio-
frequency fields – AC mains voltage
IEC 61000-4-6:2013
3 V
0.15–80 MHz
6 V
ISM frequency bands
0.15–80 MHz
80% AM at 1 kHz
Compliant
Immunity to voltage dips, short interruptions and voltage
variations
IEC 61000-4-11:2004
Compliant
Electromagnetic compatibility (EMC)
Interference immunity measurements SIP/SOP
Immunity to electrostatic discharge
IEC 61000-4-2:2008
± 8 kV contact
± 2kV, ± 4 kV, ± 8 kV, ± 15 kV air
Compliant
Immunity to electrical fast transients/bursts – I/O,
SIP/SOP ports
IEC 61000-4-4:2012
± 1 kV
100 kHz repetition rate
Compliant
Immunity to impulse voltages, conductor to earth
IEC 61000-4-5:2005
± 2 kV
Immunity to conducted disturbances, induced by radio-
frequency fields – SIP/SOP ports
IEC 61000-4-6:2013
3 V
0.15–80 MHz
6 V
ISM frequency bands
0.15–80 MHz
80% AM at 1 kHz
Compliant
Product description
16
7119100007L02 1909V001
EN