A949100002L02 1806V002
19
We recommend changing the amalgam
collector vessel when it reaches 95% full.
Disposable amalgam container filled to
100 %
– The disposable amalgam container must be
changed.
Wear protective gloves.
Use a face mask.
Wear protective goggles.
– Only when the amalgam collecting container
has been replaced is the amalgam separator
again ready for operation.
Language selection
i
Press on
language
The context menu with the list of possible
languages opens. The current language is
marked.
i
Select language and confirm with
OK
.
Select information
i
Press on
Information
.
Result:
The display shows the following information.
– Item number (REF)
– Network configuration (IP address)
– Firmware version installed on the device
– Software version on the unit
Select user authorisations
The software is equipped with two user authori-
sations:
9 Operation
9.1 Unit with display
The software interface consists of the menu bar
and the contents area.
1
2
1 Menu bar
2 Contents area
You can scroll with scroll bars if there are several
selection options.
The views and rights depend on the se-
lected operator rights (user or service
technician).
Symbols
The messages for a device are marked with the
following symbols. The symbols also show the
status of a device.
Malfunction
Notice
Note
Information
Disposable amalgam container filled to 95 %
Usage
EN
Содержание Power Tower Silence 200
Страница 1: ...EN 1806V002 A949100002L02 Installation and operating instructions Power Tower Silence 200 ...
Страница 2: ......
Страница 25: ...A949100002L02 1806V002 23 11 Connection media plan PTS 200 Appendix EN ...
Страница 26: ...24 A949100002L02 1806V002 Appendix PTS 200 PC board mit with CA 4 EN ...
Страница 27: ...A949100002L02 1806V002 25 Appendix EN ...
Страница 33: ...A949100002L02 1806V002 31 Appendix Y1 Relief valve EN ...
Страница 34: ...32 A949100002L02 1806V002 Appendix X3 Top display optional X4 Manifold control contact EN ...
Страница 38: ...36 A949100002L02 1806V002 Appendix M5 Drying unit fan Y1 Relief valve Q4 Pressure switch EN ...
Страница 39: ...A949100002L02 1806V002 37 Appendix X4 24 V AC Ablagesteuerkontakt EN ...
Страница 42: ...40 A949100002L02 1806V002 Appendix Y1 Relief valve EN ...
Страница 43: ...A949100002L02 1806V002 41 Appendix X4 24 V AC manifold control contact EN ...
Страница 46: ...44 A949100002L02 1806V002 Appendix EN ...
Страница 47: ......