The following types of hoses must not be
used:
– Rubber hoses
– Completely PVC hoses
– Hoses that are not sufficiently flexible
6.3 Installation and routeing of hos-
es and pipes
i
Execute the on-site pipe installation in accord-
ance with the applicable local regulations and
standards.
i
Lay the hose installation of the drains to or
from the unit at a sufficient incline.
If incorrectly laid, the hoses can be-
come blocked with sedimentation.
6.4 Information about electrical
connections
i
Ensure that the electrical connections to the
mains power supply are established in ac-
cordance with current valid national and local
regulations and standards governing the in-
stallation of low voltage units in medical facili-
ties.
i
The connection to the mains supply must be a
fixed connection that can only be released us-
ing a tool. Plug-inn connections (power outlet/
plug) are not permissible.
i
Install an all-pole disconnection unit (all-pole
switch) in the electrical connection to the
mains power supply. This must include the
clearance and creepage distances defined in
IEC61058-1 for a mains voltage peak of 4kV.
The disconnection unit (switch) must be
easily accessible without danger.
i
Observe the current consumption of the de-
vices that are to be connected.
Electrical fusing
LS switch 16 A, characteristic B, C and D in ac-
cordance with 60898.
Only qualified specialists or employees
trained by Dürr Dental are permitted to
install, connect and start using the unit.
6 Requirements
Further information can be found in our
suction planning information leaflet. Order
number 9000-617-03/..
The room chosen for installation must satisfy the
following requirements:
– Closed, dry, well-ventilated room
– It should not be a room made for another
purpose (e.g. boiler room or wet cell).
– Set up the unit on a clean, level and suf-
ficiently stable surface (take the weight of the
unit into account).
– Set up the units so that the type plate can be
easily read and the unit is easily accessible for
operation and maintenance.
– Leave sufficient distance from any wall (at
least 20 cm).
– Refer to the requirements for environmental
conditions in "4 Technical data".
6.1 Pipe materials
Only use vacuum-sealed HT-waste pipes
manufactured from the following materials:
– Polypropylene (PP),
– Chlorinated polyvinyl chloride (PVC-C),
– Unplasticized polyvinyl chloride (PVC-U),
– Polyethylene (PE).
The following materials must not be used:
– Acrylonitrile-butadiene-styrene (ABS),
– Styrene copolymer blends (e.g. SAN + PVC).
6.2 Hose materials
For waste connections and suction lines only
use the following hose types:
– Flexible spiral hoses made of PVC with inte-
grated spiral or equivalent hoses
– Hoses that are resistant to dental disinfectants
and chemicals
Plastic hoses will display signs of ageing
over time. Therefore, they should be in-
spected regularly and replaced as neces-
sary.
Assembly
EN
Содержание Power Tower Silence 200
Страница 1: ...EN 1806V002 A949100002L02 Installation and operating instructions Power Tower Silence 200 ...
Страница 2: ......
Страница 25: ...A949100002L02 1806V002 23 11 Connection media plan PTS 200 Appendix EN ...
Страница 26: ...24 A949100002L02 1806V002 Appendix PTS 200 PC board mit with CA 4 EN ...
Страница 27: ...A949100002L02 1806V002 25 Appendix EN ...
Страница 33: ...A949100002L02 1806V002 31 Appendix Y1 Relief valve EN ...
Страница 34: ...32 A949100002L02 1806V002 Appendix X3 Top display optional X4 Manifold control contact EN ...
Страница 38: ...36 A949100002L02 1806V002 Appendix M5 Drying unit fan Y1 Relief valve Q4 Pressure switch EN ...
Страница 39: ...A949100002L02 1806V002 37 Appendix X4 24 V AC Ablagesteuerkontakt EN ...
Страница 42: ...40 A949100002L02 1806V002 Appendix Y1 Relief valve EN ...
Страница 43: ...A949100002L02 1806V002 41 Appendix X4 24 V AC manifold control contact EN ...
Страница 46: ...44 A949100002L02 1806V002 Appendix EN ...
Страница 47: ......